<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

7 Block OUTPUTS<br />

Function description<br />

OUTPUTS CURRENT OUTPUT (1...3) CONFIGURATION<br />

FAILSAFE MODE<br />

(4006)<br />

For safety reasons it is advisable to ensure that the current output assumes a predefined<br />

state in the event <strong>of</strong> a fault. The setting you select here affects only the current output. It<br />

has no effect on other outputs and the display (e.g. totalizers).<br />

Options:<br />

MIN. CURRENT<br />

The current output adopts the value <strong>of</strong> the lower signal on alarm level (as defined in the<br />

function CURRENT SPAN (4001), see page 64.<br />

MAX. CURRENT<br />

The current output adopts the value <strong>of</strong> the upper signal on alarm level (as defined in the<br />

function CURRENT SPAN (4001), see page 64.<br />

HOLD VALUE (not recommended)<br />

Measuring value output is based on the last measuring value saved before the error<br />

occurred .<br />

ACTUAL VALUE<br />

Measured value output is based on the current flow measurement.<br />

The fault is ignored .<br />

Factory setting:<br />

MIN. CURRENT<br />

Issue/Rev. 0.2 (12/12) MN0M023 • Page 71