7 Block OUTPUTS <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

7.1 Group CURRENT OUTPUT (1...3)<br />

7.1.1 Function group CONFIGURATION<br />

OUTPUTS E CURRENT OUTPUT 1 EAA CONFIGURATION 400<br />

<br />

CURRENT OUTPUT 2 EAB CONFIGURATION 400<br />

<br />

CURRENT OUTPUT 3 EAC CONFIGURATION 400<br />

Function description<br />

OUTPUTS CURRENT OUTPUT (1...3) CONFIGURATION<br />

ASSIGN CURRENT<br />

OUTPUT<br />

(4000)<br />

Use this function to assign a measured variable to the current output.<br />

Options:<br />

OFF<br />

MASS FLOW<br />

VOLUME FLOW<br />

CORRECTED VOLUME FLOW<br />

DENSITY<br />

REFERENCE DENSITY<br />

TEMPERATURE<br />

Advanced options with optional s<strong>of</strong>tware package BATCHING:<br />

BATCH UPWARDS (batching progress upwards)<br />

BATCH DOWNWARDS (batching progress downwards)<br />

! Note!<br />

• The current range selected (function CURRENT SPAN (4001)) corresponds to a<br />

batching rate <strong>of</strong> 0 – 100%, based on the batching quantity.<br />

• The batching s<strong>of</strong>tware automatically sets the values for 0/4 mA and 20 mA (function<br />

VALUE 0_4 mA (4002) and VALUE 20 mA (4003)).<br />

Example with upward batching:<br />

Value 0/4 mA = 0 [unit]; value 20 mA = batching quantity [unit].<br />

Advanced options with optional s<strong>of</strong>tware package CONCENTRATION:<br />

TARGET MASS FLOW<br />

% TARGET MASS FLOW<br />

TARGET VOLUME FLOW<br />

% TARGET VOLUME FLOW<br />

CORRECTED TARGET VOLUME FLOW<br />

CARRIER MASS FLOW<br />

% CARRIER MASS FLOW<br />

CARRIER VOLUME FLOW<br />

% CARRIER VOLUME FLOW<br />

CORRECTED CARRIER VOLUME FLOW<br />

% BLACK LIQUOR<br />

°BAUME<br />

°API<br />

°PLATO<br />

°BALLING<br />

°BRIX<br />

OTHERS (_ _ _ _ flexible concentration)<br />

(continued on next page)<br />

Page 62 • MN0M023<br />

Issue/Rev. 0.2 (12/12)