6 Block TOTALIZERS <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

6.2 Group HANDLING TOTALIZER<br />

TOTALIZERS D TOTALIZER 1 DAA<br />

<br />

TOTALIZER 2 DAB<br />

<br />

TOTALIZER 3 DAC<br />

<br />

HANDLING TOTALIZER DJA Handling totalizer functions<br />

Function description<br />

TOTALIZERS HANDLING TOTALIZER Handling totalizer functions<br />

RESET ALL TOTALIZERS<br />

(3800)<br />

Use this function to reset the totals (including all overflows) <strong>of</strong> the totalizers (1...3) to<br />

“zero” (= RESET).<br />

Options:<br />

NO<br />

YES<br />

Factory setting:<br />

NO<br />

! Note!<br />

If the device has a status input and if it is appropriately configured, a reset for the totalizer<br />

(1...3) can also be triggered by a pulse (see the ASSIGN STATUS INPUT function (5000)<br />

on page 112).<br />

FAILSAFE ALL<br />

TOTALIZERS<br />

(3801)<br />

Use this function to define the common response <strong>of</strong> all totalizers (1...3) in case <strong>of</strong> error.<br />

Options:<br />

STOP<br />

The totalizer is paused until the fault is rectified.<br />

ACTUAL VALUE<br />

The totalizer continues to count based on the current flow measuring value. The fault is<br />

ignored .<br />

HOLD VALUE<br />

The totalizer continues to count the flow that based on the last valid flow value (before<br />

the fault occurred).<br />

Factory setting:<br />

STOP<br />

Page 60 • MN0M023<br />

Issue/Rev. 0.2 (12/12)