<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

6 Block TOTALIZERS<br />

6.1 Group TOTALIZER (1...3)<br />

6.1.1 Function group CONFIGURATION<br />

TOTALIZERS D TOTALIZER 1 DAA CONFIGURATION 300<br />

<br />

TOTALIZER 2 DAB CONFIGURATION 300<br />

<br />

TOTALIZER 3 DAC CONFIGURATION 300<br />

Function description<br />

TOTALIZERS TOTALIZER (1...3) CONFIGURATION<br />

The function descriptions below apply to totalizers 1...3; the totalizers are independently configurable.<br />

ASSIGN<br />

(3000)<br />

Use this function to assign a measured variable to the totalizer in question.<br />

Options (standard):<br />

OFF<br />

MASS FLOW<br />

VOLUME FLOW<br />

CORRECTED VOLUME FLOW<br />

Advanced options with optional s<strong>of</strong>tware package CONCENTRATION:<br />

TARGET MASS<br />

TARGET VOLUME<br />

TARGET CORRECTED VOLUME<br />

CARRIER MASS<br />

CARRIER VOLUME<br />

CARRIER CORRECTED VOLUME<br />

Factory setting:<br />

MASS FLOW<br />

! Note!<br />

• The totalizer is reset to “0” as soon as the selection is changed.<br />

• If you select OFF in the function group CONFIGURATION <strong>of</strong> the totalizer in question,<br />

only the ASSIGN (3000) function remains visible.<br />

Issue/Rev. 0.2 (12/12) MN0M023 • Page 57