17.01.2015 Views

Proline Promass 83E/F/O Description of Device Functions - FMC ...

Proline Promass 83E/F/O Description of Device Functions - FMC ...

Proline Promass 83E/F/O Description of Device Functions - FMC ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

4 Block QUICK SETUP<br />

4.3 Gas measurement Setup menu<br />

XXX.XXX.XX<br />

Esc<br />

- +<br />

E<br />

E<br />

++<br />

Quick Setup<br />

B<br />

HOME-POSITION<br />

+E<br />

QS 1004<br />

Gas measurement<br />

Assign<br />

Low flow cut <strong>of</strong>f<br />

6400<br />

On-value<br />

Low flow cut <strong>of</strong>f<br />

6402<br />

Off-value<br />

Low flow cut <strong>of</strong>f<br />

6403<br />

A0002502-en<br />

Settings for the Gas Measurement Setup menu:<br />

Fct. code Function name Suggested settings <strong>Description</strong><br />

Call up through the function matrix:<br />

B QUICK SETUP QUICK SETUP GAS MEASUR. see p. 25<br />

1004 QUICK SETUP GAS MEASUR. YES see p. 25<br />

Basic configuration:<br />

6420 EMPTY PIPE DETECTION No entry possible, the selection changes<br />

automatically to OFF.<br />

see p. 125<br />

6400 ASSIGN LOW FLOW CUT OFF For gas measurement we recommend<br />

to switch <strong>of</strong>f the low flow cut <strong>of</strong>f.<br />

OFF<br />

see p. 122<br />

6402 ON-VALUE LOW FLOW CUT<br />

OFF<br />

If you don’t switch <strong>of</strong>f the low flow cut <strong>of</strong>f:<br />

0.0000<br />

see p. 122<br />

6403 OFF-VALUE LOW FLOW CUT<br />

OFF<br />

If you don’t switch <strong>of</strong>f the low flow cut <strong>of</strong>f:<br />

50%<br />

see p. 122<br />