4 Block QUICK SETUP <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

! Note!<br />

• The display returns to the cell QUICK SETUP PULSATING FLOW (1003) if you press the<br />

X key combination during parameter interrogation.<br />

• You can call up this Setup menu either directly from the “COMMISSIONING” Setup menu or<br />

manually by means <strong>of</strong> the function QUICK SETUP PULSATING FLOW (1003).<br />

Settings for the Pulsating Flow Setup menu:<br />

Fct. code Function name Suggested settings <strong>Description</strong><br />

Call up through the function matrix:<br />

B QUICK SETUP QUICK SETUP PULSATING FLOW see p. 25<br />

1003 QUICK SETUP PULS.FLOW YES see p. 25<br />

Basic configuration:<br />

2002 DISPLAY DAMPING 1 second see p. 36<br />

3002 TOTALIZER MODE (DAA) BALANCE see p. 58<br />

3002 TOTALIZER MODE (DAB) BALANCE see p. 58<br />

3002 TOTALIZER MODE (DAC) BALANCE see p. 58<br />

Select the signal type: CURRENT OUTPUT (1...n)<br />

4004 MEASURING MODE PULSATING FLOW see p. 68<br />

4005 TIME CONSTANT 1 second see p. 70<br />

Select the signal type: FREQ./PULSE OUTPUT (1...2) / operating mode: FREQUENCY<br />

4206 MEASURING MODE PULSATING FLOW see p. 79<br />

4208 TIME CONSTANT 0 seconds see p. 84<br />

Select the signal type: FREQ./PULSE OUTPUT (1...2) / operating mode: PULSE<br />

4225 MEASURING MODE PULSATING FLOW see p. 87<br />

Other settings:<br />

8005 ALARM DELAY 0 seconds see p. 181<br />

6400 ASSIGN LF CUT OFF MASS FLOW see p. 122<br />

6402 ON-VALUE LOW FLOW CUT<br />

OFF<br />

Depends on diameter (DN [mm]):<br />

– DN 1 = 0.02 [kg/h] or [l/h]<br />

– DN 2 = 0.10 [kg/h] or [l/h]<br />

– DN 4 = 0.45 [kg/h] or [l/h]<br />

– DN 8 = 2.0 [kg/h] or [l/h]<br />

– DN 15 = 6.5 [kg/h] or [l/h]<br />

– DN 15* = 18 [kg/h] or [l/h]<br />

– DN 25 = 18 [kg/h] or [l/h]<br />

– DN 25* = 45 [kg/h] or [l/h]<br />

– DN 40 = 45 [kg/h] or [l/h]<br />

– DN 40* = 70 [kg/h] or [l/h]<br />

– DN 50 = 70 [kg/h] or [l/h]<br />

– DN 50* = 180 [kg/h] or [l/h]<br />

– DN 80 = 180 [kg/h] or [l/h]<br />

– DN 100 = 350 [kg/h] or [l/h]<br />

– DN 150 = 650 [kg/h] or [l/h]<br />

– DN 250 = 1800 [kg/h] or [l/h]<br />

see p. 122<br />

6403 OFF-VALUE LOW FLOW CUT<br />

OFF<br />

6404 PRESSURE SHOCK<br />

SUPPRESSION<br />

*DN 15, 25, 40 “FB” = Full bore<br />

versions <strong>Promass</strong> I<br />

50% see p. 122<br />

0 s see p. 123<br />

Page 30 • MN0M023<br />

Issue/Rev. 0.2 (12/12)