4 Block QUICK SETUP <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
! Note!<br />
• The display returns to the cell QUICK SETUP PULSATING FLOW (1003) if you press the<br />
X key combination during parameter interrogation.<br />
• You can call up this Setup menu either directly from the “COMMISSIONING” Setup menu or<br />
manually by means <strong>of</strong> the function QUICK SETUP PULSATING FLOW (1003).<br />
Settings for the Pulsating Flow Setup menu:<br />
Fct. code Function name Suggested settings <strong>Description</strong><br />
Call up through the function matrix:<br />
B QUICK SETUP QUICK SETUP PULSATING FLOW see p. 25<br />
1003 QUICK SETUP PULS.FLOW YES see p. 25<br />
Basic configuration:<br />
2002 DISPLAY DAMPING 1 second see p. 36<br />
3002 TOTALIZER MODE (DAA) BALANCE see p. 58<br />
3002 TOTALIZER MODE (DAB) BALANCE see p. 58<br />
3002 TOTALIZER MODE (DAC) BALANCE see p. 58<br />
Select the signal type: CURRENT OUTPUT (1...n)<br />
4004 MEASURING MODE PULSATING FLOW see p. 68<br />
4005 TIME CONSTANT 1 second see p. 70<br />
Select the signal type: FREQ./PULSE OUTPUT (1...2) / operating mode: FREQUENCY<br />
4206 MEASURING MODE PULSATING FLOW see p. 79<br />
4208 TIME CONSTANT 0 seconds see p. 84<br />
Select the signal type: FREQ./PULSE OUTPUT (1...2) / operating mode: PULSE<br />
4225 MEASURING MODE PULSATING FLOW see p. 87<br />
Other settings:<br />
8005 ALARM DELAY 0 seconds see p. 181<br />
6400 ASSIGN LF CUT OFF MASS FLOW see p. 122<br />
6402 ON-VALUE LOW FLOW CUT<br />
OFF<br />
Depends on diameter (DN [mm]):<br />
– DN 1 = 0.02 [kg/h] or [l/h]<br />
– DN 2 = 0.10 [kg/h] or [l/h]<br />
– DN 4 = 0.45 [kg/h] or [l/h]<br />
– DN 8 = 2.0 [kg/h] or [l/h]<br />
– DN 15 = 6.5 [kg/h] or [l/h]<br />
– DN 15* = 18 [kg/h] or [l/h]<br />
– DN 25 = 18 [kg/h] or [l/h]<br />
– DN 25* = 45 [kg/h] or [l/h]<br />
– DN 40 = 45 [kg/h] or [l/h]<br />
– DN 40* = 70 [kg/h] or [l/h]<br />
– DN 50 = 70 [kg/h] or [l/h]<br />
– DN 50* = 180 [kg/h] or [l/h]<br />
– DN 80 = 180 [kg/h] or [l/h]<br />
– DN 100 = 350 [kg/h] or [l/h]<br />
– DN 150 = 650 [kg/h] or [l/h]<br />
– DN 250 = 1800 [kg/h] or [l/h]<br />
see p. 122<br />
6403 OFF-VALUE LOW FLOW CUT<br />
OFF<br />
6404 PRESSURE SHOCK<br />
SUPPRESSION<br />
*DN 15, 25, 40 “FB” = Full bore<br />
versions <strong>Promass</strong> I<br />
50% see p. 122<br />
0 s see p. 123<br />
Page 30 • MN0M023<br />
Issue/Rev. 0.2 (12/12)