<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

11 Block SUPERVISION<br />

11.1.2 Function group OPERATION<br />

SUPERVISION J SYSTEM JAA CONFIGURATION 800<br />

<br />

OPERATION 804<br />

Function description<br />

SUPERVISION SYSTEM OPERATION<br />

ACTUAL SYSTEM<br />

CONDITION<br />

(8040)<br />

Use this function to check the present system condition.<br />

Display:<br />

“SYSTEM OK” or the fault / notice message with the highest priority.<br />

PREVIOUS SYSTEM<br />

CONDITION<br />

(8041)<br />

Use this function to view the fifteen most recent fault and notice messages since<br />

measuring last started.<br />

Display:<br />

The 15 most recent fault or notice messages.<br />

SIMULATION FAILSAFE<br />

MODE<br />

(8042)<br />

Use this function to set all inputs, outputs and totalizers to their defined failsafe modes, in<br />

order to check whether they respond correctly. During this time, the message<br />

“SIMULATION FAILSAFE MODE” appears on the display.<br />

Options:<br />

ON<br />

OFF<br />

Factory setting:<br />

OFF<br />

SIMULATION<br />

MEASURAND<br />

(8043)<br />

Use this function to set all inputs, outputs and totalizers to their defined flow-response<br />

modes, in order to check whether they respond correctly. During this time, the message<br />

“SIMULATION MEASURAND” appears on the display.<br />

Options:<br />

OFF<br />

MASS FLOW<br />

VOLUME FLOW<br />

CORRECTED VOLUME FLOW<br />

DENSITY<br />

REFERENCE DENSITY<br />

TEMPERATURE<br />

Factory setting:<br />

OFF<br />

" Caution!<br />

• The measuring device cannot be used for measuring while this simulation is in<br />

progress.<br />

• The setting is not saved if the power supply fails.<br />

Issue/Rev. 0.2 (12/12) MN0M023 • Page 183