11 Block SUPERVISION <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

Function description<br />

SUPERVISION SYSTEM CONFIGURATION<br />

REMOVE<br />

SW-OPTION<br />

(8006)<br />

! Note!<br />

This function is only available if:<br />

• The F-CHIP s<strong>of</strong>tware options were saved beforehand<br />

• The F-CHIP is not located on the I/O board <strong>of</strong> the measuring device<br />

Deletes all F-CHIP s<strong>of</strong>tware options, such as batching, density functions, etc.<br />

Options:<br />

0 = NO<br />

1 = YES<br />

Factory setting: NO<br />

" Caution!<br />

If process variables which are only available via the F-CHIP s<strong>of</strong>tware options are assigned<br />

to the local display or the outputs, these have to be reconfigured. If reconfiguration does<br />

not take place, the local display and the totalizer are set to the factory setting and the<br />

outputs are set to OFF.<br />

PERMANENT STRORAGE<br />

(8007)<br />

This function indicates whether permanent storage <strong>of</strong> all parameters in the EEPROM has<br />

been switched on or <strong>of</strong>f.<br />

Display:<br />

“OFF” or “ON”<br />

Factory setting:<br />

ON<br />