11 Block SUPERVISION <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

11.1 Group SYSTEM<br />

11.1.1 Function group CONFIGURATION<br />

SUPERVISION J SYSTEM JAA CONFIGURATION 800<br />

Function description<br />

SUPERVISION SYSTEM CONFIGURATION<br />

ASSIGN SYSTEM ERROR<br />

(8000)<br />

Use this function to view all system errors.<br />

By selecting a certain system error, its error category can be changed in the subsequent<br />

function ERROR CATEGORY (8001).<br />

Options:<br />

CANCEL<br />

List <strong>of</strong> system errors<br />

! Note!<br />

• You can exit this function as follows: select “CANCEL” and confirm with F.<br />

• A list <strong>of</strong> possible system errors is provided in the Operating Instruction<br />

Operating Instructions <strong>Proline</strong> <strong>Promass</strong> 83, BA 059D/06/en/<br />

ERROR CATEGORY<br />

(8001) ! Note!<br />

This function is only available if a system error has been selected in the function ASSIGN<br />

SYSTEM ERROR (8000).<br />

Use this function to define whether a system error triggers a notice message or a fault<br />

message. If you select FAULT MESSAGES, all outputs respond to an error in accordance<br />

with their defined error response patterns.<br />

Options:<br />

NOTICE MESSAGES (display only)<br />

FAULT MESSAGES (outputs and display)<br />

! Note!<br />

Press the key F twice to call up the function ASSIGN SYSTEM ERROR (8000).<br />

ASSIGN PROCESS ERROR<br />

(8002)<br />

Use this function to view all process errors.<br />

By selecting an individual process error, its error category can be changed in the<br />

subsequent function ERROR CATEGORY (8003).<br />

Options:<br />

CANCEL<br />

List <strong>of</strong> process errors<br />

! Note!<br />

• You can exit this function as follows: select “CANCEL” and confirm with F.<br />

• A list <strong>of</strong> possible process errors is provided in the Operating Instruction<br />