10 Block SPECIAL FUNCTION <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
10.3.7 Function group TUBE DAMPING<br />
SPECIAL FUNCTION H ADVANCED DIAGNOSIS HEA CONFIGURATION 740<br />
<br />
ACQUISITION 741<br />
<br />
MASS FLOW 742<br />
<br />
DENSITY 743<br />
<br />
REFERENCE DENSITY 744<br />
<br />
TEMPERATURE 745<br />
<br />
TUBE DAMPING 746<br />
Function description<br />
SPECIAL FUNCTION ADVANCED DIAGNOSIS TUBE DAMPING<br />
REFERENCE VALUE<br />
TUBE DAMPING<br />
(7460)<br />
TUBE DAMPING<br />
(7461)<br />
MINIMUM VALUE<br />
(7462)<br />
The reference value for tube damping appears on the display.<br />
Display:<br />
5-digit floating-point number<br />
The measured tube damping appears on the display.<br />
Display:<br />
5-digit floating-point number<br />
The lowest tube damping value since the saved values were last reset appears on the<br />
display.<br />
Display:<br />
5-digit floating-point number<br />
MAXIMUM VALUE<br />
(7463)<br />
The highest tube damping value since the saved values were last reset appears on the<br />
display.<br />
Display:<br />
5-digit floating-point number<br />
TUBE DAMPING<br />
HISTORY<br />
(7464)<br />
TUBE DAMPING<br />
DEVIATION<br />
(7465)<br />
The last ten tube damping values since the saved values were last reset appears on the<br />
display.<br />
Display:<br />
5-digit floating-point number<br />
This function displays the deviation between the measured tube damping and the<br />
reference values (FACTORY or USER), see page 166, selected in the function SELECT<br />
REFERENCE CONDITION (7402).<br />
Display:<br />
5-digit floating-point number<br />
WARNING LEVEL<br />
(7466) ! Note!<br />
This function is not available unless ON was selected in the WARNING MODE function<br />
(7403).<br />
Use this function to specify a limit value for tube damping. A notice message is generated<br />
if the limit value is exceeded.<br />
User input:<br />
0...99999 [%]<br />
Factory setting:<br />
1000%<br />
Page 172 • MN0M023<br />
Issue/Rev. 0.2 (12/12)