<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
10 Block SPECIAL FUNCTION<br />
10.3.6 Function group TEMPERATURE<br />
SPECIAL FUNCTION H ADVANCED DIAGNOSIS HEA CONFIGURATION 740<br />
<br />
ACQUISITION 741<br />
<br />
MASS FLOW 742<br />
<br />
DENSITY 743<br />
<br />
REFERENCE DENSITY 744<br />
<br />
TEMPERATURE 745<br />
Function description<br />
SPECIAL FUNCTION ADVANCED DIAGNOSIS TEMPERATURE<br />
! Note!<br />
The appropriate unit is taken from the function UNIT TEMPERATURE (0422) (see page 21).<br />
REFERENCE VALUE<br />
TEMPERATURE<br />
(7450)<br />
TEMPERATURE<br />
(7451)<br />
MINIMUM VALUE<br />
(7452)<br />
The reference value for the temperature appears on the display.<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
The currently measured temperature appears on the display.<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
The lowest temperature value since the saved values were last reset appears on the<br />
display.<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
MAXIMUM VALUE<br />
(7453)<br />
The highest temperature value since the saved values were last reset appears on the<br />
display.<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
TEMPERATURE HISTORY<br />
(7454)<br />
The last ten temperature values since the saved values were last reset appear on the<br />
display.<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
TEMPERATURE<br />
DEVIATION<br />
(7455)<br />
This function displays the deviation between the currently measured temperature and the<br />
reference values (FACTORY or USER), see page 166, selected in the function SELECT<br />
REFERENCE CONDITION (7402).<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
WARNING LEVEL<br />
(7456) ! Note!<br />
This function is not available unless ON was selected in the WARNING MODE function<br />
(7403).<br />
Use this function to specify a limit value for the temperature. A notice message is<br />
generated if the limit value is exceeded.<br />
User input:<br />
0...99999 °C<br />
Factory setting:<br />
100 °C<br />
Issue/Rev. 0.2 (12/12) MN0M023 • Page 171