10 Block SPECIAL FUNCTION <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

10.3.5 Function group REFERENCE DENSITY<br />

SPECIAL FUNCTION H ADVANCED DIAGNOSIS HEA CONFIGURATION 740<br />

<br />

ACQUISITION 741<br />

<br />

MASS FLOW 742<br />

<br />

DENSITY 743<br />

<br />

REFERENCE DENSITY 744<br />

Function description<br />

SPECIAL FUNCTION ADVANCED DIAGNOSIS REFERENCE DENSITY<br />

! Note!<br />

The appropriate unit is taken from the function UNIT REFERENCE DENSITY (0421) ( page 20).<br />

REFERENCE VALUE<br />

REFERENCE DENSITY<br />

(7440)<br />

The reference value for the reference density appears on the display.<br />

Display:<br />

5-digit floating-point number, incl. units<br />

REFERENCE DENSITY<br />

(7441)<br />

The measured reference density appears on the display.<br />

Display:<br />

5-digit floating-point number, incl. units<br />

MINIMUM VALUE<br />

(7442)<br />

The lowest reference density value since the saved values were last reset appears on the<br />

display.<br />

Display:<br />

5-digit floating-point number, incl. units<br />

MAXIMUM VALUE<br />

(7443)<br />

The highest reference density value since the saved values were last reset appears on the<br />

display.<br />

Display:<br />

5-digit floating-point number, incl. units<br />

REFERENCE DENSITY<br />

HISTORY<br />

(7444)<br />

REFERENCE DENSITY<br />

DEVIATION<br />

(7445)<br />

The last ten reference density values since the saved values were last reset appear on the<br />

display.<br />

Display:<br />

5-digit floating-point number, incl. units<br />

This function displays the deviation between the measured reference density and the<br />

reference values (FACTORY or USER), see page 166, selected in the function SELECT<br />

REFERENCE CONDITION (7402).<br />

Display:<br />

5-digit floating-point number, incl. units<br />

WARNING LEVEL<br />

(7446) ! Note!<br />

This function is not available unless ON was selected in the WARNING MODE function<br />

(7403).<br />

Use this function to specify a limit value for the reference density. A notice message is<br />

generated if the limit value is exceeded.<br />

User input:<br />

0...99999 [%]<br />

Factory setting:<br />

100%<br />

Page 170 • MN0M023<br />