Proline Promass 83E/F/O Description of Device Functions - FMC ...
Proline Promass 83E/F/O Description of Device Functions - FMC ...
Proline Promass 83E/F/O Description of Device Functions - FMC ...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
10 Block SPECIAL FUNCTION <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
10.3.3 Function group MASS FLOW<br />
SPECIAL FUNCTION H ADVANCED DIAGNOSIS HEA CONFIGURATION 740<br />
<br />
ACQUISITION 741<br />
<br />
MASS FLOW 742<br />
Function description<br />
SPECIAL FUNCTION ADVANCED DIAGNOSIS MASS FLOW<br />
! Note!<br />
The appropriate unit is taken from the function UNIT MASS FLOW (0400) (see page 17).<br />
REFERENCE VALUES<br />
MASS FLOW<br />
(7420)<br />
The reference value for the mass flow appears on the display.<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
MASS FLOW<br />
(7421)<br />
The measured mass flow appears on the display.<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
MINIMUM VALUE<br />
(7422)<br />
The lowest mass flow value since the saved values were last reset appears on the display.<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
MAXIMUM VALUE<br />
(7423)<br />
The highest mass flow value since the saved values were last reset appears on the display.<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
MASS FLOW HISTORY<br />
(7424)<br />
The last ten mass flow values since the saved values were last reset appear on the display.<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
MASS FLOW DEVIATION<br />
(7425)<br />
This function displays deviation between the measured mass flow and the reference<br />
values (FACTORY or USER), see page 166, selected in the function SELECT REFERENCE<br />
CONDITION (7402).<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
WARNING LEVEL<br />
(7426) ! Note!<br />
This function is not available unless ON was selected in the WARNING MODE function<br />
(7403).<br />
Use this function to specify a limit value for the mass flow. A notice message is generated<br />
if the limit value is exceeded.<br />
User input:<br />
0...99999 [Mass flow unit]<br />
Factory setting:<br />
90000 kg/h<br />
Page 168 • MN0M023<br />
Issue/Rev. 0.2 (12/12)