10 Block SPECIAL FUNCTION<br />
10.3.2 Function group ACQUISITION<br />
SPECIAL FUNCTION H ADVANCED DIAGNOSIS HEA CONFIGURATION 740<br />
<br />
ACQUISITION 741<br />
Function description<br />
SPECIAL FUNCTION ADVANCED DIAGNOSIS ACQUISITION<br />
ACQUISITION MODE<br />
(7410)<br />
Use this function to specify whether the advanced diagnosis parameters should be<br />
determined on a periodical or single-shot basis.<br />
Options:<br />
OFF<br />
PERIODICAL<br />
SINGLE SHOT<br />
Factory setting:<br />
OFF<br />
! Note!<br />
See the Chapter on “Commissioning” in the Operating Instructions <strong>Proline</strong> <strong>Promass</strong> 83,<br />
BA 059D/06/en for more information on advanced diagnosis.<br />
ACQUISITION PERIOD<br />
(7411) ! Note!<br />
This function is not available unless PERIODICAL was selected in the ACQUISITION<br />
MODE function (7410).<br />
Use this function to specify the time interval after which the advanced diagnosis<br />
parameters should be taken. The time interval starts with the confirmation <strong>of</strong> the input.<br />
User input:<br />
0...99999 s<br />
Factory setting:<br />
3600 s<br />
! Note!<br />
A reference status must be defined prior to determining the diagnosis parameters,<br />
see function SELECT REFERENCE CONDITION (7402).<br />
ACQUISITION DO<br />
(7412) ! Note!<br />
This function is not available unless SINGLE SHOT was selected in the ACQUISITION<br />
MODE function (7410).<br />
Use this function to start determining the advanced diagnosis parameters on a single-shot<br />
basis.<br />
Options:<br />
START – CANCEL<br />
Factory setting:<br />
CANCEL<br />
! Note!<br />
A reference status must be defined prior to determining the diagnosis parameters,<br />
see function SELECT REFERENCE CONDITION (7402).<br />
RESET HISTORY<br />
(7413)<br />
Use this function to delete all history values.<br />
Options:<br />
YES – NO<br />
Factory setting:<br />
NO<br />
Issue/Rev. 0.2 (12/12) MN0M023 • Page 167