<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

10 Block SPECIAL FUNCTION<br />

10.3.2 Function group ACQUISITION<br />

SPECIAL FUNCTION H ADVANCED DIAGNOSIS HEA CONFIGURATION 740<br />

<br />

ACQUISITION 741<br />

Function description<br />

SPECIAL FUNCTION ADVANCED DIAGNOSIS ACQUISITION<br />

ACQUISITION MODE<br />

(7410)<br />

Use this function to specify whether the advanced diagnosis parameters should be<br />

determined on a periodical or single-shot basis.<br />

Options:<br />

OFF<br />

PERIODICAL<br />

SINGLE SHOT<br />

Factory setting:<br />

OFF<br />

! Note!<br />

See the Chapter on “Commissioning” in the Operating Instructions <strong>Proline</strong> <strong>Promass</strong> 83,<br />

BA 059D/06/en for more information on advanced diagnosis.<br />

ACQUISITION PERIOD<br />

(7411) ! Note!<br />

This function is not available unless PERIODICAL was selected in the ACQUISITION<br />

MODE function (7410).<br />

Use this function to specify the time interval after which the advanced diagnosis<br />

parameters should be taken. The time interval starts with the confirmation <strong>of</strong> the input.<br />

User input:<br />

0...99999 s<br />

Factory setting:<br />

3600 s<br />

! Note!<br />

A reference status must be defined prior to determining the diagnosis parameters,<br />

see function SELECT REFERENCE CONDITION (7402).<br />

ACQUISITION DO<br />

(7412) ! Note!<br />

This function is not available unless SINGLE SHOT was selected in the ACQUISITION<br />

MODE function (7410).<br />

Use this function to start determining the advanced diagnosis parameters on a single-shot<br />

basis.<br />

Options:<br />

START – CANCEL<br />

Factory setting:<br />

CANCEL<br />

! Note!<br />

A reference status must be defined prior to determining the diagnosis parameters,<br />

see function SELECT REFERENCE CONDITION (7402).<br />

RESET HISTORY<br />

(7413)<br />

Use this function to delete all history values.<br />

Options:<br />

YES – NO<br />

Factory setting:<br />

NO<br />