10 Block SPECIAL FUNCTION <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

10.3 Group ADVANCED DIAGNOSIS<br />

10.3.1 Function group CONFIGURATION<br />

SPECIAL FUNCTION H ADVANCED DIAGNOSIS HEA CONFIGURATION 740<br />

Function description<br />

SPECIAL FUNCTION ADVANCED DIAGNOSIS CONFIGURATION<br />

REFERENCE<br />

CONDITION USER<br />

(7401)<br />

Use this function to start determining the user reference status. The following values are<br />

determined:<br />

• MASS FLOW<br />

• DENSITY<br />

• REFERENCE DENSITY<br />

• TEMPERATURE<br />

• TUBE DAMPING<br />

• ELECTRODYNAMIC SENSORS<br />

• OPERATING FREQUENCY FLUCTUATION<br />

• TUBE DAMPING FLUCTUATION<br />

Options:<br />

CANCEL<br />

START<br />

Factory setting:<br />

CANCEL<br />

SELECT REFERENCE<br />

CONDITION<br />

(7402)<br />

Use this function to select the reference status which should be used to compare the<br />

advanced diagnosis parameters (see function ACQUISITION MODE (7410) on<br />

page 167).<br />

Options:<br />

FACTORY<br />

USER<br />

Factory setting:<br />

FACTORY<br />

WARNING MODE<br />

(7403)<br />

Use this function to determine whether a warning should be generated when there is a<br />

deviation between the reference status (FACTORY or USER, see function SELECT<br />

REFERENCE CONDITION (7402)) and the current measuring values.<br />

The values <strong>of</strong> the following functions are compared to the reference status :<br />

• MASS FLOW (7421)<br />

• DENSITY (7431)<br />

• REFERENCE DENSITY (7441)<br />

• TEMPERATURE (7451)<br />

• TUBE DAMPING (7461)<br />

• ELECTRODYNAMIC SENSORS (7471)<br />

• OPERATING FREQUENCY FLUCTUATION (7481)<br />

• TUBE DAMPING FLUCTUATION (7491)<br />

Options:<br />

OFF<br />

ON<br />

Factory setting:<br />

OFF<br />

Page 166 • MN0M023<br />

Issue/Rev. 0.2 (12/12)