10.2.5 Function group OPERATION<br />

SPECIAL FUNCTION H BATCHING FUNCTION HCA CONFIGURATION 720<br />

<br />

VALVE PARAMETER 722<br />

<br />

SUPERVISION 724<br />

<br />

OPERATION 726<br />

Function description<br />

SPECIAL FUNCTION BATCHING FUNCTION OPERATION<br />

BATCH PROCEDURE<br />

(7260)<br />

Use this function to control a batching process. The batching can be started manually or a<br />

batching already running can be interrupted or stopped at any time.<br />

Options:<br />

STOP (Stop batching)<br />

START (Start batching)<br />

HOLD (Interrupt batching)<br />

GO ON (Continue batching)<br />

Factory setting:<br />

STOP<br />

! Note!<br />

• This function can also be controlled via the status input, (see function ASSIGN<br />

STATUS INPUT (5000) on page 112).<br />

• If the information line has been assigned to BATCHING MENU (see page 50), the<br />

application-specific functions <strong>of</strong> the minus key (START-STOP) and the plus key<br />

(HOLD-GO ON / batching specification) are defined locally. In this way, a direct<br />

batching control station is available locally at the measuring device by means <strong>of</strong> the<br />

user interface (not access-protected).<br />

• In the event <strong>of</strong> a fault:<br />

– during the batching process, the batching is canceled (STOP) and the local display<br />

alternates between displaying the batching menu and the fault message.<br />

• If the positive zero return is activated, (see page 133):<br />

– during the batching process, the batching is canceled (STOP).<br />

– during a pause in the batching (option HOLD), the batching cannot be restarted,<br />

(see also notice messages # 571 and # 572 in the Operating Instructions <strong>Proline</strong><br />

<strong>Promass</strong> 83, BA 059D/06/en, Chapter Troubleshooting.<br />

BATCH UPWARDS<br />

(7261)<br />

In this function the batching progress can be read upwards, i.e. starting at 0 the quantity<br />

displayed increases until the batching process is complete.<br />

Display:<br />

Floating-point number incl. unit<br />

! Note!<br />

The value <strong>of</strong> this function can be output via the current output.<br />