<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

10 Block SPECIAL FUNCTION<br />

Function description<br />

SPECIAL FUNCTION BATCHING FUNCTION SUPERVISION<br />

MAX. FLOW<br />

(7244)<br />

A maximum flow value can be specified in this function. The batching process is aborted<br />

and all the valves are closed if the specified flow value is overshot.<br />

Application:<br />

This function can be used for safety reasons, for example, to ensure all batching valves<br />

close in the event <strong>of</strong> a system fault.<br />

User input:<br />

5-digit floating-point number<br />

Factory setting:<br />

0 [unit] (= deactivated)<br />

! Note!<br />

• The appropriate unit is taken depending on the process variable selected in the<br />

ASSIGN BATCH VARIABLE parameter and the unit configured in the SYSTEM UNITS<br />

function group.<br />

• The function is not active if you enter 0 (factory setting).<br />

• If the batching process is aborted because the specified flow value is overshot, the<br />

parameter BATCH COUNTER is not incremented.<br />

• New error message > MAX. FLOW with the error number #474. The error message is<br />

output automatically after 60 s.<br />

In the function ASSIGN PROCESS ERROR (8002), you can use the ERROR CATEGORY<br />

(8003) to define whether this should be treated as a fault or notice message.<br />

Factory setting = FAULT MESSAGE<br />

Issue/Rev. 0.2 (12/12) MN0M023 • Page 161