<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
10 Block SPECIAL FUNCTION<br />
Function description<br />
SPECIAL FUNCTION BATCHING FUNCTION SUPERVISION<br />
MAX. FLOW<br />
(7244)<br />
A maximum flow value can be specified in this function. The batching process is aborted<br />
and all the valves are closed if the specified flow value is overshot.<br />
Application:<br />
This function can be used for safety reasons, for example, to ensure all batching valves<br />
close in the event <strong>of</strong> a system fault.<br />
User input:<br />
5-digit floating-point number<br />
Factory setting:<br />
0 [unit] (= deactivated)<br />
! Note!<br />
• The appropriate unit is taken depending on the process variable selected in the<br />
ASSIGN BATCH VARIABLE parameter and the unit configured in the SYSTEM UNITS<br />
function group.<br />
• The function is not active if you enter 0 (factory setting).<br />
• If the batching process is aborted because the specified flow value is overshot, the<br />
parameter BATCH COUNTER is not incremented.<br />
• New error message > MAX. FLOW with the error number #474. The error message is<br />
output automatically after 60 s.<br />
In the function ASSIGN PROCESS ERROR (8002), you can use the ERROR CATEGORY<br />
(8003) to define whether this should be treated as a fault or notice message.<br />
Factory setting = FAULT MESSAGE<br />
Issue/Rev. 0.2 (12/12) MN0M023 • Page 161