10 Block SPECIAL FUNCTION<br />

Function description<br />

SPECIAL FUNCTION BATCHING FUNCTION SUPERVISION<br />

MIN. BATCHING<br />

QUANTITY<br />

(7241)<br />

Use this function to specify a minimum batching quantity. A message is generated if the<br />

minimum batching quantity was not achieved by the time batching ends (e.g. if after run<br />

mode is active). The quantity value is entered as a % or as an absolute value, depending<br />

on the option in the function INPUT FORMAT (7209).<br />

Application:<br />

Message stating that underbatching is present (e.g. the contents <strong>of</strong> the containers does<br />

not correspond to the quantity declared).<br />

User input:<br />

0 to max. value or 0 to 100% (related to the batch quantity)<br />

Factory setting:<br />

0 [unit] (= deactivated)<br />

" Caution!<br />

• When the batching quantity is adjusted (reduced/increased), (see function BATCH<br />

QUANTITY (7203) on page 148), there is no automatic adjustment, i.e. this value<br />

must be determined again and reentered (see also the fault message # 472 in the<br />

Operating Instructions <strong>Proline</strong> <strong>Promass</strong> 83, BA 059D/06/en, Troubleshooting<br />

Chapter).<br />

• Batching (START) is not possible when the fault message is active!<br />

! Note!<br />

• The function is not active if you enter 0 (factory setting).<br />

• As a factory setting, this function is assigned a fault message. This appears for<br />

60 seconds. The fault message can be acknowledged earlier by changing a batching<br />

function.<br />

• If this function is intended more for general monitoring purposes or if there is a short<br />

time lapse between two batching processes, it is advisable to assign this function to a<br />

notice message, (see function ERROR CATEGORY on page 181). However, while the<br />

notice message is active (60 seconds) the next batching can be started and the notice<br />

message is acknowledged.<br />

• This function can be output via the switch output.<br />

Issue/Rev. 0.2 (12/12) MN0M023 • Page 159