<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

10 Block SPECIAL FUNCTION<br />

Function description<br />

SPECIAL FUNCTION BATCHING FUNCTION CONFIGURATION<br />

CALCULATION MODE<br />

(7206) ! Note!<br />

This function is only available if MODE 1 or MODE 2 was selected in the<br />

COMPENSATION MODE function (7206).<br />

Use this function to select the method for calculating the recorded after run quantities.<br />

Options:<br />

ALL<br />

All after run quantities are used in the calculation.<br />

SELECTION<br />

The after run quantities recorded are filtered. The smallest and largest after run quantity<br />

are not taken into account in the calculation (extreme value filter).<br />

Factory setting:<br />

ALL<br />

! Note!<br />

Machine-related (larger) “extreme values”, occurring especially at startup, delay<br />

correction and distort real reproducibility. By selecting “SELECT”, these “extreme<br />

values” are not taken into account.<br />

Example:<br />

Function CALCULATION MODE (7206) = SELECTION<br />

Function AVERAGING DRIP (7207) = 5<br />

From five after run quantities recorded, the largest and smallest are not used. From the<br />

remaining three after run quantities, an average after run quantity is calculated which is<br />

taken into account at the next batching.<br />

AVERAGING DRIP<br />

(7207) ! Note!<br />

This function is only available if MODE 1 or MODE 2 was selected in the<br />

COMPENSATION MODE function (7206).<br />

Use this function to specify the number <strong>of</strong> after run quantities (cycles) which are included<br />

in calculating the batching compensation mode, MODE 1 or MODE 2.<br />

! Note!<br />

The value entered in this function influences the measuring system reaction time.<br />

If you specify:<br />

• a small calculation depth (low value entered) = measuring system reacts quickly to<br />

differing amount <strong>of</strong> after run quantities.<br />

• a large calculation depth (high value entered) = measuring system reacts slowly to<br />

changing after run quantities.<br />

User input:<br />

0...100<br />

Factory setting:<br />

0 [cycles]<br />

Issue/Rev. 0.2 (12/12) MN0M023 • Page 151