10 Block SPECIAL FUNCTION <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

Function description<br />

SPECIAL FUNCTION BATCHING FUNCTION CONFIGURATION<br />

ASSIGN BATCH<br />

VARIABLE<br />

(7202)<br />

Use this function to assign a batching variable to the batching specification.<br />

Options:<br />

OFF<br />

MASS FLOW<br />

VOLUME FLOW<br />

CORRECTED VOLUME FLOW<br />

Advanced options with optional s<strong>of</strong>tware package CONCENTRATION:<br />

TARGET MASS<br />

TARGET VOLUME<br />

TARGET CORRECTED VOLUME<br />

CARRIER MASS<br />

CARRIER VOLUME<br />

CARRIER CORRECTED VOLUME<br />

Factory setting:<br />

OFF<br />

! Note!<br />

• The possible assignments <strong>of</strong> the display functions are automatically extended. Once a<br />

batching variable has been selected (MASS or VOLUME), you can locally define the<br />

application-specific function <strong>of</strong> the minus key (start-stop-continue) and the plus key<br />

(stop-batching name/quantity) in the information line by means <strong>of</strong> the “batching<br />

menu” assignment. In this way, a direct batching control station is made available<br />

locally at the measuring device by means <strong>of</strong> the user interface and the controls.<br />

• Select OFF if the BATCHING functionality is no longer to be used. All settings related<br />

to the function (e.g. switching contact assigned to the relay output) must be assigned<br />

to another functionality.<br />

BATCH QUANTITY<br />

(7203)<br />

Use this function to define the quantity to be batched.<br />

User input:<br />

5-digit floating-point number<br />

Factory setting:<br />

0 [unit]<br />

! Note!<br />

• The appropriate unit is taken from the function group SYSTEM UNITS (ACA),<br />

(see page 17).<br />

• When the batching quantity entered here is achieved, valve 1 closes (see function<br />

CLOSE VALVE 1 (7221) on page 153).<br />

Page 148 • MN0M023<br />

Issue/Rev. 0.2 (12/12)