10 Block SPECIAL FUNCTION <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
Function description<br />
SPECIAL FUNCTION BATCHING FUNCTION CONFIGURATION<br />
ASSIGN BATCH<br />
VARIABLE<br />
(7202)<br />
Use this function to assign a batching variable to the batching specification.<br />
Options:<br />
OFF<br />
MASS FLOW<br />
VOLUME FLOW<br />
CORRECTED VOLUME FLOW<br />
Advanced options with optional s<strong>of</strong>tware package CONCENTRATION:<br />
TARGET MASS<br />
TARGET VOLUME<br />
TARGET CORRECTED VOLUME<br />
CARRIER MASS<br />
CARRIER VOLUME<br />
CARRIER CORRECTED VOLUME<br />
Factory setting:<br />
OFF<br />
! Note!<br />
• The possible assignments <strong>of</strong> the display functions are automatically extended. Once a<br />
batching variable has been selected (MASS or VOLUME), you can locally define the<br />
application-specific function <strong>of</strong> the minus key (start-stop-continue) and the plus key<br />
(stop-batching name/quantity) in the information line by means <strong>of</strong> the “batching<br />
menu” assignment. In this way, a direct batching control station is made available<br />
locally at the measuring device by means <strong>of</strong> the user interface and the controls.<br />
• Select OFF if the BATCHING functionality is no longer to be used. All settings related<br />
to the function (e.g. switching contact assigned to the relay output) must be assigned<br />
to another functionality.<br />
BATCH QUANTITY<br />
(7203)<br />
Use this function to define the quantity to be batched.<br />
User input:<br />
5-digit floating-point number<br />
Factory setting:<br />
0 [unit]<br />
! Note!<br />
• The appropriate unit is taken from the function group SYSTEM UNITS (ACA),<br />
(see page 17).<br />
• When the batching quantity entered here is achieved, valve 1 closes (see function<br />
CLOSE VALVE 1 (7221) on page 153).<br />
Page 148 • MN0M023<br />
Issue/Rev. 0.2 (12/12)