<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
10 Block SPECIAL FUNCTION<br />
Function description<br />
SPECIAL FUNCTION DENSITY FUNCTIONS CONFIGURATION<br />
SQR. EXPANSION<br />
COEFFICIENT TARGET<br />
FLUID<br />
(7006)<br />
! Note!<br />
This function is not available unless one <strong>of</strong> the following was selected in the DENSITY<br />
FUNCTION function (7000):<br />
• % MASS / % VOLUME<br />
• %-BLACK LIQUOR<br />
Use this function to enter the fluid-specific square expansion coefficient for the carrier<br />
fluid for nonlinear temperature curves. This value is required for temperaturecompensated<br />
calculation <strong>of</strong> the target fluid content in a<br />
two-phase fluid.<br />
User input:<br />
5-digit floating-point number, including unit and sign<br />
Factory setting:<br />
0.0000 e–6 [1/K 2 ]<br />
LINEAR EXPANSION<br />
COEFFICIENT<br />
(7007)<br />
! Note!<br />
This function is not available unless one <strong>of</strong> the following was selected in the DENSITY<br />
FUNCTION function (7000):<br />
• °BAUME < 1 SG<br />
• °BAUME > 1 SG<br />
•°API<br />
• °PLATO<br />
• °BALLING<br />
• °BRIX<br />
Use this function to enter the fluid-specific linear expansion coefficient (for linear<br />
temperature curves), to calculate the temperature-compensated density functions.<br />
User input:<br />
5-digit floating-point number, including unit and sign<br />
Factory setting:<br />
0.5000 e–3 [1/K]<br />
SQR. EXPANSION<br />
COEFFICIENT<br />
(7008)<br />
! Note!<br />
This function is not available unless one <strong>of</strong> the following was selected in the DENSITY<br />
FUNCTION function (7000):<br />
• °BAUME < 1 SG<br />
• °BAUME > 1 SG<br />
•°API<br />
• °PLATO<br />
• °BALLING<br />
• °BRIX<br />
Use this function to enter the fluid-specific square expansion coefficient (for nonlinear<br />
temperature curves), to calculate the temperature-compensated density functions.<br />
User input:<br />
5-digit floating-point number, including unit and sign<br />
Factory setting:<br />
0.0000 e–6 [1/K 2 ]<br />
Issue/Rev. 0.2 (12/12) MN0M023 • Page 143