17.01.2015 Views

Proline Promass 83E/F/O Description of Device Functions - FMC ...

Proline Promass 83E/F/O Description of Device Functions - FMC ...

Proline Promass 83E/F/O Description of Device Functions - FMC ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

10 Block SPECIAL FUNCTION<br />

Function description<br />

SPECIAL FUNCTION DENSITY FUNCTIONS CONFIGURATION<br />

SQR. EXPANSION<br />

COEFFICIENT TARGET<br />

FLUID<br />

(7006)<br />

! Note!<br />

This function is not available unless one <strong>of</strong> the following was selected in the DENSITY<br />

FUNCTION function (7000):<br />

• % MASS / % VOLUME<br />

• %-BLACK LIQUOR<br />

Use this function to enter the fluid-specific square expansion coefficient for the carrier<br />

fluid for nonlinear temperature curves. This value is required for temperaturecompensated<br />

calculation <strong>of</strong> the target fluid content in a<br />

two-phase fluid.<br />

User input:<br />

5-digit floating-point number, including unit and sign<br />

Factory setting:<br />

0.0000 e–6 [1/K 2 ]<br />

LINEAR EXPANSION<br />

COEFFICIENT<br />

(7007)<br />

! Note!<br />

This function is not available unless one <strong>of</strong> the following was selected in the DENSITY<br />

FUNCTION function (7000):<br />

• °BAUME < 1 SG<br />

• °BAUME > 1 SG<br />

•°API<br />

• °PLATO<br />

• °BALLING<br />

• °BRIX<br />

Use this function to enter the fluid-specific linear expansion coefficient (for linear<br />

temperature curves), to calculate the temperature-compensated density functions.<br />

User input:<br />

5-digit floating-point number, including unit and sign<br />

Factory setting:<br />

0.5000 e–3 [1/K]<br />

SQR. EXPANSION<br />

COEFFICIENT<br />

(7008)<br />

! Note!<br />

This function is not available unless one <strong>of</strong> the following was selected in the DENSITY<br />

FUNCTION function (7000):<br />

• °BAUME < 1 SG<br />

• °BAUME > 1 SG<br />

•°API<br />

• °PLATO<br />

• °BALLING<br />

• °BRIX<br />

Use this function to enter the fluid-specific square expansion coefficient (for nonlinear<br />

temperature curves), to calculate the temperature-compensated density functions.<br />

User input:<br />

5-digit floating-point number, including unit and sign<br />

Factory setting:<br />

0.0000 e–6 [1/K 2 ]<br />

Issue/Rev. 0.2 (12/12) MN0M023 • Page 143

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!