10 Block SPECIAL FUNCTION<br />
10.1 Group DENSITY FUNCTIONS<br />
10.1.1 Function group CONFIGURATION<br />
SPECIAL FUNCTION H DENSITY FUNCTIONS HAA CONFIGURATION 700<br />
Function description<br />
SPECIAL FUNCTION DENSITY FUNCTIONS CONFIGURATION<br />
DENSITY FUNCTION<br />
(7000)<br />
Use this function to select the desired density function which is used to calculate special<br />
density values or the percentage proportion <strong>of</strong> components in two-phase fluids.<br />
User input:<br />
OFF<br />
% MASS / % VOLUME<br />
%-BLACK LIQUOR<br />
°BAUME > 1 SG<br />
°BAUME < 1 SG<br />
°API<br />
°PLATO<br />
°BALLING<br />
°BRIX<br />
FLEXIBLE<br />
Factory setting:<br />
OFF<br />
REFERENCE DENSITY<br />
CARRIER FLUID<br />
(7001)<br />
! Note!<br />
This function is not available unless one <strong>of</strong> the following was selected in the DENSITY<br />
FUNCTION function (7000):<br />
• % MASS / % VOLUME<br />
• %-BLACK LIQUOR<br />
Use this function to enter the reference density (density at reference temp.) <strong>of</strong> the carrier<br />
fluid. This value is required for temperature-compensated calculation <strong>of</strong> the target fluid<br />
content in a<br />
two-phase fluid.<br />
User input:<br />
5-digit floating-point number, incl. units<br />
Factory setting:<br />
1.0000 kg/l<br />
! Note!<br />
• Carrier fluid = transporting liquid, (e.g. water)<br />
Target fluid = material transported (e.g. lime powder)<br />
• The appropriate unit is taken from the function UNIT REFERENCE DENSITY (0421)<br />
(see page 20).<br />
EXPANSION<br />
COEFFICIENT LINEAR<br />
CARRIER FLUID<br />
(7002)<br />
! Note!<br />
This function is not available unless one <strong>of</strong> the following was selected in the DENSITY<br />
FUNCTION function (7000):<br />
• % MASS / % VOLUME<br />
• %-BLACK LIQUOR<br />
Use this function to enter the fluid-specific linear expansion coefficient for the carrier<br />
fluid for linear temperature curves. This value is required for temperature-compensated<br />
calculation <strong>of</strong> the target fluid content in a<br />
two-phase fluid.<br />
User input:<br />
5-digit floating-point number, including unit and sign<br />
Factory setting:<br />
0.5000 e –3 [1/K]<br />
Issue/Rev. 0.2 (12/12) MN0M023 • Page 141