3 Block MEASURED VARIABLES <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
Function description<br />
MEASURED VARIABLES MEASURING VALUES ADDITIONAL VALUES<br />
CORRECTED TARGET<br />
VOLUME FLOW<br />
(0024)<br />
! Note!<br />
This function is not available unless % MASS / % VOLUME was selected in the DENSITY<br />
FUNCTION function (7000), (see page 141).<br />
Use this function to display the currently measured corrected volume flow <strong>of</strong> the target<br />
fluid. Target fluid = carried material (e.g. lime powder).<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
CARRIER MASS FLOW<br />
(0025) ! Note!<br />
This function is not available unless one <strong>of</strong> the following was selected:<br />
• in the function DENSITY FUNCTION (7000), see page 141:<br />
– % MASS / % VOLUME<br />
– FLEXIBLE and in the function MODE (7010), see page 144, the selection<br />
% MASS 2D or % MASS 3D<br />
Use this function to display the currently measured mass flow <strong>of</strong> the carrier fluid.<br />
Carrier fluid = transporting liquid (e.g. water).<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
% CARRIER MASS FLOW<br />
(0026) ! Note!<br />
This function is not available unless one <strong>of</strong> the following was selected:<br />
• in the function DENSITY FUNCTION (7000), see page 141:<br />
– % MASS / % VOLUME<br />
– FLEXIBLE and in the function MODE (7010), see page 144, the selection<br />
% MASS 2D or % MASS 3D<br />
Use this function to display the currently measured mass flow <strong>of</strong> the carrier fluid as a %<br />
(<strong>of</strong> the overall mass flow). Carrier fluid = transporting liquid (e.g. water).<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
CARRIER VOLUME<br />
FLOW<br />
(0027)<br />
! Note!<br />
This function is not available unless one <strong>of</strong> the following was selected:<br />
• in the function DENSITY FUNCTION (7000), see page 141:<br />
– % MASS / % VOLUME<br />
– FLEXIBLE and in the function MODE (7010), see page 144, the selection<br />
% VOLUME 2D or % VOLUME 3D<br />
Use this function to display the currently measured volume flow <strong>of</strong> the carrier fluid.<br />
Carrier fluid = transporting liquid (e.g. water).<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
Page 14 • MN0M023<br />
Issue/Rev. 0.2 (12/12)