<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
9 Block BASIC FUNCTION<br />
9.4 Group SENSOR DATA<br />
9.4.1 Function group CONFIGURATION<br />
BASIC FUNCTION G HART GAA<br />
<br />
PROCESS PARAMETER GIA<br />
<br />
SYSTEM PARAMETER GLA<br />
<br />
SENSOR DATA GNA CONFIGURATION 680<br />
Function description<br />
BASIC FUNCTION SENSOR DATA CONFIGURATION<br />
All sensor data (calibration factor, zero point and nominal diameter) are set at the factory and saved on the S-DAT sensor<br />
memory chip.<br />
" Caution!<br />
Under normal circumstances you should not change the following parameter settings, because changes affect numerous<br />
functions <strong>of</strong> the entire measuring facility in general and the accuracy <strong>of</strong> the measuring system in particular. For this<br />
reason, the functions described below cannot be changed even when you enter your personal code.<br />
Contact the E+H service organization if you have any questions about these functions.<br />
! Note!<br />
The individual values <strong>of</strong> the functions are also provided on the sensor nameplate.<br />
K-FACTOR<br />
(6800)<br />
This function shows the current calibration factor for the sensor.<br />
Factory setting:<br />
Depends on nominal diameter and calibration<br />
ZEROPOINT<br />
(6803)<br />
This function shows the current zero point correction value for the sensor.<br />
Display:<br />
max. 5-digit number: -99999...+99999<br />
Factory setting:<br />
Depends on calibration<br />
NOMINAL DIAMETER<br />
(6804)<br />
This function shows the nominal diameter for the sensor.<br />
Factory setting:<br />
Depends on the size <strong>of</strong> the sensor<br />
Issue/Rev. 0.2 (12/12) MN0M023 • Page 135