9 Block BASIC FUNCTION<br />

9.3 Group SYSTEM PARAMETER<br />

9.3.1 Function group CONFIGURATION<br />

BASIC FUNCTION G HART GAA<br />

<br />

PROCESS PARAMETER GIA<br />

<br />

SYSTEM PARAMETER GLA CONFIGURATION 660<br />

Function description<br />

BASIC FUNCTION SYSTEM PARAMETER CONFIGURATION<br />

INSTALLATION<br />

DIRECTION SENSOR<br />

(6600)<br />

Use this function to reverse the sign <strong>of</strong> the flow direction, if necessary.<br />

! Note!<br />

Ascertain the actual direction <strong>of</strong> fluid flow with reference to the direction indicated by<br />

the arrow on the sensor (nameplate).<br />

Options:<br />

NORMAL (flow as indicated by the arrow)<br />

INVERSE (flow opposite to direction indicated by the arrow)<br />

Factory setting:<br />

NORMAL<br />

DENSITY DAMPING<br />

(6602)<br />

The density filter allows the sensitivity <strong>of</strong> the density measuring signal to be lowered with<br />

respect to variations in the density <strong>of</strong> the fluid, e.g. with inhomogeneous liquids. The<br />

damping acts on all functions and outputs <strong>of</strong> the measuring device.<br />

User input:<br />

max. 5-digit number, including unit: 0.00...100.00 s<br />

Factory setting:<br />

0.00 s<br />

FLOW DAMPING<br />

(6603)<br />

Setting the filter depth <strong>of</strong> the digital filter. The sensitivity <strong>of</strong> the flow measurement signal<br />

can be reduced with respect to interference peaks (e.g. in the event <strong>of</strong> a high solid<br />

content, gas bubbles in the fluid etc.). The reaction time <strong>of</strong> the measuring device<br />

increases with every increase in the filter setting. The damping acts on all functions and<br />

outputs <strong>of</strong> the measuring device.<br />

User input:<br />

0...100 s<br />

Factory setting:<br />

Liquid: 0,00 s<br />

Gas: 0,25 s<br />

POSITIVE ZERO RETURN<br />

(6605)<br />

Use this function to interrupt evaluation <strong>of</strong> measured variables. This is necessary when a<br />

piping system is being cleaned, for example.<br />

This setting acts on all function and outputs <strong>of</strong> the measuring device.<br />

Options:<br />

OFF<br />

ON (signal output is set to the “ZERO FLOW” value, temperature and density are still<br />

output)<br />

Factory setting:<br />

OFF<br />

Issue/Rev. 0.2 (12/12) MN0M023 • Page 133