9 Block BASIC FUNCTION <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
Function description<br />
BASIC FUNCTION PROCESS PARAMETER ADJUSTMENT<br />
DENSITY ADJUST MODE<br />
(6482)<br />
Use this function to select whether a 1-point or a 2-point density adjustment should be<br />
carried out.<br />
Options:<br />
CANCEL<br />
1-POINT<br />
2-POINT<br />
DENSITY SETPOINT 1<br />
(6483)<br />
Use this function to enter the density setpoint value for the first fluid for which you want<br />
to carry out field density adjustment.<br />
User input:<br />
5-digit floating-point number, incl. units<br />
! Note!<br />
• The preset density entered here should not vary from the actual fluid density by a more<br />
than ±10%.<br />
• The appropriate unit is taken from the function group SYSTEM UNITS (see page 17).<br />
MEASURE FLUID 1<br />
(6484)<br />
In this function the actual density <strong>of</strong> the first fluid is measured for density adjustment.<br />
Options:<br />
CANCEL<br />
START<br />
DENSITY SETPOINT 2<br />
(6485)<br />
Use this function to enter the density setpoint value for the second fluid for which you<br />
want to carry out field density adjustment.<br />
User input:<br />
5-digit floating-point number, incl. units<br />
! Note!<br />
• The preset density entered here should not vary from the actual fluid density by a more<br />
than ±10%.<br />
• The difference between the density setpoint values must be at least 0.2 kg/l.<br />
• The appropriate unit is taken from the function group SYSTEM UNITS (see page 17).<br />
MEASURE FLUID 2<br />
(6486)<br />
In this function the current density <strong>of</strong> the second fluid is measured for density<br />
adjustment.<br />
Options:<br />
CANCEL<br />
START<br />
Page 130 • MN0M023<br />
Issue/Rev. 0.2 (12/12)