9 Block BASIC FUNCTION <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

Function description<br />

BASIC FUNCTION PROCESS PARAMETER ADJUSTMENT<br />

DENSITY ADJUST MODE<br />

(6482)<br />

Use this function to select whether a 1-point or a 2-point density adjustment should be<br />

carried out.<br />

Options:<br />

CANCEL<br />

1-POINT<br />

2-POINT<br />

DENSITY SETPOINT 1<br />

(6483)<br />

Use this function to enter the density setpoint value for the first fluid for which you want<br />

to carry out field density adjustment.<br />

User input:<br />

5-digit floating-point number, incl. units<br />

! Note!<br />

• The preset density entered here should not vary from the actual fluid density by a more<br />

than ±10%.<br />

• The appropriate unit is taken from the function group SYSTEM UNITS (see page 17).<br />

MEASURE FLUID 1<br />

(6484)<br />

In this function the actual density <strong>of</strong> the first fluid is measured for density adjustment.<br />

Options:<br />

CANCEL<br />

START<br />

DENSITY SETPOINT 2<br />

(6485)<br />

Use this function to enter the density setpoint value for the second fluid for which you<br />

want to carry out field density adjustment.<br />

User input:<br />

5-digit floating-point number, incl. units<br />

! Note!<br />

• The preset density entered here should not vary from the actual fluid density by a more<br />

than ±10%.<br />

• The difference between the density setpoint values must be at least 0.2 kg/l.<br />

• The appropriate unit is taken from the function group SYSTEM UNITS (see page 17).<br />

MEASURE FLUID 2<br />

(6486)<br />

In this function the current density <strong>of</strong> the second fluid is measured for density<br />

adjustment.<br />

Options:<br />

CANCEL<br />

START<br />

Page 130 • MN0M023<br />

Issue/Rev. 0.2 (12/12)