3 Block MEASURED VARIABLES<br />
3.1.2 Function group ADDITIONAL VALUES<br />
MEASURED<br />
VARIABLES<br />
A MEASURING VALUES AAA MAIN VALUES 000<br />
<br />
ADDITIONAL VALUES 002<br />
Function description<br />
MEASURED VARIABLES MEASURING VALUES ADDITIONAL VALUES<br />
TARGET MASS FLOW<br />
(0020) ! Note!<br />
This function is not available unless one <strong>of</strong> the following was selected:<br />
• in the function DENSITY FUNCTION (7000), see page 141:<br />
– % MASS / % VOLUME<br />
– FLEXIBLE and in the function MODE (7010), see page 144, the selection<br />
% MASS 2D or % MASS 3D<br />
Use this function to display the currently measured mass flow <strong>of</strong> the target fluid. Target<br />
fluid = carried material (e.g. lime powder).<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
% TARGET MASS FLOW<br />
(0021) ! Note!<br />
This function is not available unless one <strong>of</strong> the following was selected:<br />
• in the function DENSITY FUNCTION (7000), see page 141:<br />
– % MASS / % VOLUME<br />
– FLEXIBLE and in the function MODE (7010), see page 144, the selection<br />
% MASS 2D or % MASS 3D<br />
In this function, the currently measured mass flow <strong>of</strong> the target fluid is displayed as a %<br />
(<strong>of</strong> the overall mass flow). Target fluid = carried material (e.g. lime powder).<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
TARGET VOLUME FLOW<br />
(0022) ! Note!<br />
This function is not available unless one <strong>of</strong> the following was selected:<br />
• in the function DENSITY FUNCTION (7000), see page 141:<br />
– % MASS / % VOLUME<br />
– FLEXIBLE and in the function MODE (7010), see page 144, the selection<br />
% VOLUME 2D or % VOLUME 3D<br />
In this function, the currently measured volume flow <strong>of</strong> the target fluid is displayed.<br />
Target fluid = carried material (e.g. lime powder).<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
% TARGET VOLUME<br />
FLOW<br />
(0023)<br />
! Note!<br />
This function is not available unless one <strong>of</strong> the following was selected:<br />
• in the function DENSITY FUNCTION (7000), see page 141:<br />
– % MASS / % VOLUME<br />
– FLEXIBLE and in the function MODE (7010), see page 144, the selection<br />
% VOLUME 2D or % VOLUME 3D<br />
Use this function to display the currently measured volume flow <strong>of</strong> the target fluid as a %<br />
(<strong>of</strong> the overall volume flow). Target fluid = carried material (e.g. lime powder).<br />
Display:<br />
5-digit floating-point number, including unit and sign<br />
Issue/Rev. 0.2 (12/12) MN0M023 • Page 13