<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
8 Block INPUTS<br />
8.2 Group CURRENT INPUT<br />
8.2.1 Function group CONFIGURATION<br />
INPUTS F STATUS INPUT FAA<br />
<br />
CURRENT INPUT FCA CONFIGURATION 520<br />
Function description<br />
INPUTS CURRENT INPUT CONFIGURATION<br />
ASSIGN CURRENT<br />
INPUT<br />
(5200)<br />
Use this function to assign a process variable to the current input.<br />
Options:<br />
OFF<br />
TEMPERATURE<br />
PRESSURE<br />
REFERENCE DENSITY<br />
Factory setting:<br />
OFF<br />
CURRENT SPAN<br />
(5201)<br />
Use this function to define the current range. The selection specifies the operational range<br />
and the lower and upper signal on alarm.<br />
Options:<br />
0–20 mA<br />
4–20 mA<br />
4–20 mA NAMUR<br />
4–20 mA US<br />
0–20 mA (25 mA)<br />
4–20 mA (25 mA)<br />
Factory setting:<br />
4–20 mA NAMUR<br />
! Note!<br />
When switching the hardware from an active (factory setting) to a passive output signal,<br />
select a current range <strong>of</strong> 4–20 mA (See Operating Instructions <strong>Proline</strong> <strong>Promass</strong> 83, BA<br />
059D/06/en/).<br />
Current range / operational range (measuring information):<br />
0–20 mA / 0...20.5 mA<br />
4–20 mA / 4...20.5 mA<br />
4–20 mA NAMUR / 3.8...20.5 mA<br />
4–20 mA US / 3.9...20.8 mA<br />
0–20 mA (25 mA) / 0...24 ma<br />
4–20 mA (25 mA) / 4...24 mA<br />
VALUE 0_4 mA<br />
(5202)<br />
Use this function to assign a value to the 0/4 mA current.<br />
User input:<br />
5-digit floating-point number<br />
Factory setting:<br />
Dependent on the process variable assigned to the current input (see function ASSIGN<br />
CURRENT INPUT, 5200).<br />
– Pressure: 0 barg<br />
– Temperature: –50 °C<br />
– Reference density: 0.50 kg/Nl<br />
! Note!<br />
The appropriate unit is taken from the functions UNIT PRESSURE (0426), UNIT<br />
TEMPERATURE (0422) or UNIT REFERENCE DENSITY (0421).<br />
Issue/Rev. 0.2 (12/12) MN0M023 • Page 115