8 Block INPUTS<br />
8.1.2 Function group OPERATION<br />
INPUTS F STATUS INPUT FAA CONFIGURATION 500<br />
<br />
OPERATION 504<br />
Function description<br />
INPUTS STATUS INPUT OPERATION<br />
ACTUAL<br />
STATUS INPUT<br />
(5040)<br />
Use this function to view the current level <strong>of</strong> the status input.<br />
Display:<br />
HIGH<br />
LOW<br />
SIMULATION STATUS<br />
INPUT<br />
(5041)<br />
Use this function to simulate the status input, in other words to trigger the function<br />
assigned to the status input (see the function ASSIGN STATUS INPUT (5000) on<br />
page 112).<br />
Options:<br />
OFF<br />
ON<br />
Factory setting:<br />
OFF<br />
! Note!<br />
• The “SIMULATION STATUS INPUT” message indicates that simulation is active.<br />
• The measuring device continues to measure while simulation is in progress, i.e. the<br />
current measured values are output correctly via the other outputs.<br />
" Caution!<br />
The setting is not saved if the power supply fails.<br />
VALUE SIMULATION<br />
STATUS INPUT<br />
(5042)<br />
! Note!<br />
The function is not visible unless the function SIMULATION STATUS INPUT (5041) is<br />
active (= ON).<br />
Use this function to define the level to be assumed at the status output during the<br />
simulation. This value is used to test downstream devices and the measuring device itself.<br />
Options:<br />
HIGH<br />
LOW<br />
Factory setting:<br />
LOW<br />
" Caution!<br />
The setting is not saved if the power supply fails.<br />
Issue/Rev. 0.2 (12/12) MN0M023 • Page 113