8 Block INPUTS <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
8.1 Group STATUS INPUT<br />
8.1.1 Function group CONFIGURATION<br />
INPUTS F STATUS INPUT FAA CONFIGURATION 500<br />
Function description<br />
INPUTS STATUS INPUT CONFIGURATION<br />
ASSIGN STATUS INPUT<br />
(5000)<br />
Use this function to assign a switching function to the status input.<br />
Options:<br />
OFF<br />
RESET TOTALIZER 1<br />
RESET TOTALIZER 2<br />
RESET TOTALIZER 3<br />
RESET ALL TOTALIZERS<br />
POSITIVE ZERO RETURN<br />
RESET FAULT MESSAGE<br />
ZEROPOINT ADJUSTMENT<br />
Advanced options with optional s<strong>of</strong>tware package BATCHING:<br />
RUN BATCHING (start/stop)<br />
HOLD BATCHING (stop/continue)<br />
RESET BATCH SUM (resetting total quantity / total quantity totalizers)<br />
RESET TOTALIZER 3 & START BATCHING (reset <strong>of</strong> totalizer 3, followed by start)<br />
" Caution!<br />
If an input pulse is sent during a batching process, the batching process is canceled<br />
immediately, but totalizer 3 is not reset. This allows the partial filling to be read out<br />
correctly.<br />
Advanced options with optional s<strong>of</strong>tware package ADVANCED DIAGNOSIS:<br />
ACQUISITION<br />
! Note!<br />
This function is not available unless the SINGLE SHOT setting is selected in the<br />
ACQUISITION MODE function (7410).<br />
Factory setting: OFF<br />
" Caution!<br />
POSITIVE ZERO RETURN is active as long as the level is available at the status input<br />
(continuous signal). All other assignments react to a change in level (pulse) at the status<br />
input.<br />
ACTIVE LEVEL<br />
(5001)<br />
Use this function to define whether the assigned function (see function ASSIGN STATUS<br />
INPUT) is released when the signal level is present (HIGH) or not present (LOW).<br />
Options:<br />
HIGH<br />
LOW<br />
Factory setting: HIGH<br />
MINIMUM<br />
PULSE WIDTH<br />
(5002)<br />
Use this function to define a minimum pulse width which the input pulse must achieve in<br />
order to trigger the selected switching function, (see function ASSIGN STATUS INPUT<br />
(5000)).<br />
User input:<br />
20...100 ms<br />
Factory setting: 50 ms<br />
Page 112 • MN0M023<br />
Issue/Rev. 0.2 (12/12)