<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
7 Block OUTPUTS<br />
Function description<br />
OUTPUTS RELAY OUTPUT (1...2) CONFIGURATION<br />
SWITCH-ON DELAY<br />
(4702) ! Note!<br />
This function is not available unless LIMIT VALUE or FLOW DIRECTION was selected in<br />
the ASSIGN RELAY function (4700).<br />
Use this function to define a delay (0 ... 100 seconds) for pull-up (i.e. signal changes from<br />
0 to 1) <strong>of</strong> the relay output. The delay starts when the limit value is reached. The relay<br />
output does switch when the delay has timed out and the switch on condition has been<br />
valid over the delay time.<br />
User input:<br />
fixed-point number 0.0...100.0 s<br />
Factory setting:<br />
0.0 s<br />
OFF-VALUE<br />
(4703) ! Note!<br />
This function is not available unless LIMIT VALUE was selected in the ASSIGN RELAY<br />
function (4700).<br />
Use this function to assign a value to the switch-<strong>of</strong>f point (relay drops out). The value can<br />
be higher or lower than the switch-on point. Positive or negative values are permissible,<br />
depending on the measured variable in question (e.g. mass flow, totalizer reading).<br />
User input:<br />
5-digit floating-point number [unit]<br />
Factory setting:<br />
0 [kg/h] or 2 [kg/l] or 200 [°C]<br />
! Note!<br />
• The appropriate unit is taken from the function UNIT VOLUME FLOW (0402) or<br />
UNIT MASS FLOW (0400).<br />
• If SYMMETRY is selected in the function MEASURING MODE (4705) and values with<br />
different signs are entered for the switch-on and switch-<strong>of</strong>f points, the notice message<br />
“INPUT RANGE EXCEEDED” appears.<br />
SWITCH-OFF DELAY<br />
(4704) ! Note!<br />
This function is not available unless LIMIT VALUE was selected in the ASSIGN RELAY<br />
function (4700).<br />
Use this function to define a delay (0 ... 100 seconds) for dropout (i.e. signal changes<br />
from 1 to 0) <strong>of</strong> the relay output. The delay starts when the limit value is reached.<br />
The relay output does switch when the delay has timed out and the switch <strong>of</strong>f condition<br />
has been valid over the delay time.<br />
User input:<br />
fixed-point number 0.0...100.0 s<br />
Factory setting:<br />
0.0 s<br />
Issue/Rev. 0.2 (12/12) MN0M023 • Page 103