7 Block OUTPUTS <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
Function description<br />
OUTPUTS RELAY OUTPUT (1...2) CONFIGURATION<br />
ASSIGN RELAY<br />
(continued)<br />
Advanced options with optional s<strong>of</strong>tware package CONCENTRATION:<br />
LIMIT TARGET MASS FLOW<br />
LIMIT TARGET % MASS PROPORTION FLOW<br />
LIMIT TARGET VOLUME FLOW<br />
LIMIT TARGET % VOLUME PROPORTION FLOW<br />
LIMIT TARGET CORRECTED VOLUME FLOW<br />
LIMIT CARRIER MASS FLOW<br />
LIMIT CARRIER % MASS PROPORTION FLOW<br />
LIMIT CARRIER VOLUME FLOW<br />
LIMIT CARRIER % VOLUME PROPORTION FLOW<br />
LIMIT CARRIER CORRECTED VOLUME FLOW<br />
LIMIT % BLACK LIQUOR<br />
LIMIT °BAUME > 1<br />
LIMIT °BAUME < 1<br />
LIMIT °API<br />
LIMIT °PLATO<br />
LIMIT °BALLING<br />
LIMIT °BRIX<br />
LIMIT OTHERS (_ _ _ _ flexible concentration)<br />
Advanced options with opt. s<strong>of</strong>tware package ADVANCED DIAGNOSIS:<br />
LIMIT MASS FLOW DEVIATION<br />
LIMIT DENSITY DEVIATION<br />
LIMIT REFERENCE DENSITY DEVIATION<br />
LIMIT TEMPERATURE DEVIATION<br />
LIMIT TUBE DAMPING DEVIATION<br />
LIMIT ELECTRODYNAMIC SENSOR DEVIATION<br />
LIMIT OPERATING FREQUENCY DEVIATION<br />
LIMIT TUBE DAMPING DEVIATION<br />
Factory setting:<br />
FAULT MESSAGE<br />
! Note!<br />
• It is very important to read and comply with the information on the switching<br />
characteristics <strong>of</strong> the relay output, (see page 109).<br />
• It is advisable to configure at least one relay output as a fault output and define the<br />
outputs' failsafe mode.<br />
• The relay output is configured as a normally open (NO or make) contact by default. It<br />
can be reconfigured as a normally closed (NC or break) contact by means <strong>of</strong> a jumper<br />
on the relay module (see Operating Instructions <strong>Proline</strong> <strong>Promass</strong> 83, BA<br />
059D/06/en).<br />
• If you select OFF or ON, the only function shown in the CONFIGURATION function<br />
group is this function ASSIGN RELAY (4700).<br />
ON VALUE<br />
(4701) ! Note!<br />
This function is not available unless LIMIT VALUE or FLOW DIRECTION was selected in<br />
the ASSIGN RELAY function (4700).<br />
Use this function to assign a value to the switch-on point (relay output pulls up). The<br />
value can be higher or lower than the switch-<strong>of</strong>f point. Positive or negative values are<br />
permissible, depending on the measured variable in question (e.g. mass flow, totalizer<br />
reading).<br />
User input:<br />
5-digit floating-point number [unit]<br />
Factory setting:<br />
0 [kg/h] or 2 [kg/l] or 200 [°C]<br />
! Note!<br />
• The appropriate unit is taken from the function UNIT VOLUME FLOW (0402) or<br />
UNIT MASS FLOW (0400).<br />
• Only the switch-on point is available for flow direction output (no switch-<strong>of</strong>f point). If<br />
you enter a value not equal to the zero flow (e.g. 5), the difference between the zero<br />
flow and the value entered corresponds to half the switchover hysteresis.<br />