HITACHI 704<br />

INSTRUCCIONES DE TRABAJO<br />

DIRECTIONS FOR USE<br />

AUTOMATIZACIÓN<br />

AUTOMATION<br />

<strong>LABKIT</strong> – <strong>CLINICAL</strong> <strong>DIAGNOSTIC</strong> <strong>REAGENT</strong><br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

ALBUMINE<br />

Preparation of working reagent:<br />

The reagent is ready to use.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

ALB<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 31 - 0<br />

SAMPLE VOLUME (µl) 3<br />

R1 VOLUME (µl) 500 - 50 - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 700 - 600<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

g/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 3.5 - 5<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 6 g/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

UREA UV-GLDH<br />

Preparation of working reagent:<br />

Dissolve one vial of lyophilised R2 with a little quantity of buffer R1. Return to the original vial of buffer<br />

when it is completely dissolved. The working reagent is stable 28 days refrigerated and keeps it out of<br />

light.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

TRIG.<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 31 - 0<br />

SAMPLE VOLUME (µl) 4<br />

R1 VOLUME (µl) 400 - [...] - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 600 - 505<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

mg/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 36 - 165<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 1000 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

TRIGLYCERIDES<br />

Preparation of working reagent:<br />

Dissolve one vial of lyophilised R2 with a little quantity of buffer R1. Return to the original vial of buffer<br />

when it is completely dissolved. The working reagent is stable 28 days refrigerated and keeps it out of<br />

light.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

TRIG.<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 31 - 0<br />

SAMPLE VOLUME (µl) 4<br />

R1 VOLUME (µl) 400 - [...] - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 600 - 505<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

mg/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 36 - 165<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 1000 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

TOTAL PROTEINS<br />

Preparation of working reagent:<br />

The reagent is ready to use.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

TPRO<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 31 - 0<br />

SAMPLE VOLUME (µl) 5<br />

R1 VOLUME (µl) 500 - 50 - 0<br />

R2 VOLUME (µl) 0 - 50 - 0<br />

WAVELENGHT 660 - 456<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

g/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 6.7 - 8.7<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 15 g/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

TOTAL BILIRUBIN<br />

Preparation of working reagent:<br />

Mix 25 ml. R1 with 0.5 ml. sodium nitrite. Mix well. The working reagent is stable 3 weeks to 2 - 8ºC.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

BIL-T<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 31 - 0<br />

SAMPLE VOLUME (µl) 15<br />

R1 VOLUME (µl) 350 - 20 - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 660 - 546<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

g/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 0.0 - 1.1<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 30 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

RF Turbidymetric<br />

Preparation of working reagent:<br />

Follow instructions given on the procedure.<br />

Calibration:<br />

Use RF CALIBRATOR.<br />

Operation conditions:<br />

Non-Linear Calibration.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

RF<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE [2 POINTS] - 20 - 32<br />

SAMPLE VOLUME (µl) 3<br />

R1 VOLUME (µl)<br />

350 - 20 - NO<br />

R2 VOLUME (µl)<br />

40 - 20 - NO<br />

WAVELENGHT 600<br />

CALIB. METHOD NON-LINEAR 1 - 5<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. 16.5 - 2<br />

STD 3 CONC. RACK POS. 33 – 3<br />

STD 4 CONC. RACK POS. 66 – 4<br />

STD 5 CONC. RACK POS. 132 – 5<br />

STD 6 CONC. RACK POS. 0.0 – 6<br />

UNIT<br />

IU/ml.<br />

SD LIMIT 300<br />

DUPLICATE LIMIT 1000<br />

SENSITIVITY 0<br />

ABS. LIMIT<br />

0 - INCREASE<br />

PROZONE LIMIT<br />

0 - LOWER<br />

EXPECTED VALUE 0 - 20<br />

INSTRUMENT FACTOR 1.0<br />

*Depends on the concentration of calibrated used.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

PHOSPHORUS<br />

Preparation of working reagent:<br />

The reagents are ready to use.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

PHOSPH<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 32 – 0<br />

SAMPLE VOLUME (µl) 7<br />

R1 VOLUME (µl) 350 - 50 - 0<br />

R2 VOLUME (µl) 0 - 50 - 0<br />

WAVELENGHT 376 - 340<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

mg/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 2.5 - 5.0<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 15 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

PHOSPHOLIPIDES<br />

Preparation of working reagent:<br />

Dissolve 1 vial lyophilised R2 with 10 ml. buffer R1. Homogenise. The working reagent is stable 2 weeks<br />

refrigerated and keeps it out of the light.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

PHLPD<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 31 - 0<br />

SAMPLE VOLUME (µl) 4<br />

R1 VOLUME (µl) 350 - 50 - 0<br />

R2 VOLUME (µl) 0 - 50 - 0<br />

WAVELENGHT 600 - 505<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - [...]<br />

UNIT<br />

mg/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 160 - 270<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 600 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

MAGNESIUM<br />

Preparation of working reagent:<br />

The reagent is ready to use.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

MG<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 2 - 15 - 32<br />

SAMPLE VOLUME (µl) 5<br />

R1 VOLUME (µl) 200 - 100 - 0<br />

R2 VOLUME (µl) 200 - 100 - 0<br />

WAVELENGHT 600 - 505<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

mg/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 1.6 - 2.5<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 6 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

Lp(a) Turbidymetric<br />

Preparation of working reagent:<br />

Follow instructions given on the procedure.<br />

Calibration:<br />

Use Lp(a) CALIBRATOR.<br />

Operation conditions:<br />

Non-Linear Calibration.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

Lp(a)<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE [2 POINTS] - 20 - 32<br />

SAMPLE VOLUME (µl) 3<br />

R1 VOLUME (µl)<br />

350 - 20 - NO<br />

R2 VOLUME (µl)<br />

40 - 20 - NO<br />

WAVELENGHT 600<br />

CALIB. METHOD NON-LINEAR 1 - 5<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. 100 - 2<br />

STD 3 CONC. RACK POS. 200 - 3<br />

STD 4 CONC. RACK POS. 400 - 4<br />

STD 5 CONC. RACK POS. 800 - 5<br />

STD 6 CONC. RACK POS. 0.0 - 6<br />

UNIT<br />

mg/ml.<br />

SD LIMIT 300<br />

DUPLICATE LIMIT 1000<br />

SENSITIVITY 0<br />

ABS. LIMIT<br />

0 - INCREASE<br />

PROZONE LIMIT<br />

0 - LOWER<br />

EXPECTED VALUE 0 - 300<br />

INSTRUMENT FACTOR 1.0<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

LDH<br />

Preparation of working reagent:<br />

Dissolve one tablet with 15 ml. of buffer. The working reagent is stable 3 days refrigerated.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

LDH<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 5 - 5 - 12<br />

SAMPLE VOLUME (µl) 10<br />

R1 VOLUME (µl) 350 - [...] - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 376 - 340<br />

CALIB. METHOD 2 - 0<br />

STD 1 CONC. RACK POS. 0 – 1<br />

UNIT<br />

U/L.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 5000 - 1<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 0 - 460<br />

INSTRUMENT FACTOR 1.00<br />

3-CALIBRATION MONITOR<br />

K=-8571<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 1500 U/L.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

LACTATE<br />

Preparation of working reagent:<br />

Dissolve 1 vial lyophilised R2 with 10 ml of buffer R1. Homogenise. The working reagent is stable 4 weeks<br />

2-8ºC. and keep it out of the light.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

LACT<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 31 – 0<br />

SAMPLE VOLUME (µl) 4<br />

R1 VOLUME (µl) 350 - 20 - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 600 - 505<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - [...]<br />

UNIT<br />

mg/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 9.0 - 16.0<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 150 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

HDL-CHOLESTEROL<br />

Preparation of working reagent:<br />

Precipitated reagent.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Make separation of HDL fraction, following the procedure instructions.<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

HDL<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 31 - 0<br />

SAMPLE VOLUME (µl) 10<br />

R1 VOLUME (µl) 300 - 20 - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 700 - 505<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - [...]<br />

UNIT<br />

mg/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 30 - 65<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 300 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

GPT / ALT<br />

Preparation of working reagent:<br />

Dissolve one tablet into one vial of buffer or one tablet with 15 ml. of buffer, depending of presentation.<br />

The working reagent is stable 2 weeks refrigerated.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

GPT<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 5 - 5 - 15<br />

SAMPLE VOLUME (µl) 20<br />

R1 VOLUME (µl) 350 - 50 - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 376 - 340<br />

CALIB. METHOD 2 - 0<br />

STD 1 CONC. RACK POS. 0 – 1<br />

UNIT<br />

U/L.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 5000 - 1<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 0 - 40<br />

INSTRUMENT FACTOR 1.00<br />

3-CALIBRATION MONITOR<br />

K=3500<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 500 U/L.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

GOT / AST<br />

Preparation of working reagent:<br />

Dissolve one tablet into one vial of buffer or one tablet with 15 ml. of buffer, depending of presentation.<br />

The working reagent is stable 2 weeks refrigerated.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

GOT<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 5 - 5 - 15<br />

SAMPLE VOLUME (µl) 20<br />

R1 VOLUME (µl) 350 - 50 - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 376 - 340<br />

CALIB. METHOD 2 - 0<br />

STD 1 CONC. RACK POS. 0 – 1<br />

UNIT<br />

U/L.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 5000 - 1<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 0 - 35<br />

INSTRUMENT FACTOR 1.00<br />

3-CALIBRATION MONITOR<br />

K=3500<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 500 U/L.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

GLUCOSE Hexokinase<br />

Preparation of working reagent:<br />

Dissolve 1 vial lyophilised R2 with a little quantity of buffer R1. Return to the original vial of buffer when it is<br />

completely dissolved. The working reagent is stable 1 month refrigerated and keep it out of the light.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

GLU-H<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 31 – 0<br />

SAMPLE VOLUME (µl) 3<br />

R1 VOLUME (µl) 300 - 100 - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 405 - 340<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

mg/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

PANIC VALUE [...] - [...]<br />

EXPECTED VALUE 60 - 110<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 700 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

GLUCOSE<br />

Preparation of working reagent:<br />

Dissolve 1 vial lyophilised R2 with a little quantity of buffer R1. Return to the original vial of buffer when it is<br />

completely dissolved. The working reagent is stable 1 month refrigerated and keep it out of the light.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

GLU<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 31 – 0<br />

SAMPLE VOLUME (µl) 4<br />

R1 VOLUME (µl) 400 - 50 - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 600 - 505<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

mg/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 60 - 110<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 500 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

γ-GT<br />

Preparation of working reagent:<br />

Dissolve one tablet into one vial of buffer or one tablet with 15 ml. of buffer, depending of presentation.<br />

The working reagent is stable 21 days refrigerated.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

GGT<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 5 - 5 - 12<br />

SAMPLE VOLUME (µl) 20<br />

R1 VOLUME (µl) 350 - 50 - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 660 - 415<br />

CALIB. METHOD 2 - 0<br />

STD 1 CONC. RACK POS. 0 – 1<br />

UNIT<br />

U/L.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 15000 - 1<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 0 - 50<br />

INSTRUMENT FACTOR 1.00<br />

3-CALIBRATION MONITOR<br />

K=3000<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 500 U/L.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

FERRITIN Turbidymetric<br />

Preparation of working reagent:<br />

Follow instructions given on the procedure.<br />

Calibration:<br />

Use FERRITIN CALIBRATOR.<br />

Operation conditions:<br />

Non-Linear Calibration.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

FERRITIN<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE [2 POINTS] - 19 - 32<br />

SAMPLE VOLUME (µl) 20<br />

R1 VOLUME (µl)<br />

230 - 20 - NO<br />

R2 VOLUME (µl)<br />

85 - 20 - NO<br />

WAVELENGHT 600<br />

CALIB. METHOD NON-LINEAR 1 - 5<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. 87.4 - 2<br />

STD 3 CONC. RACK POS. 160 - 3<br />

STD 4 CONC. RACK POS. 314 - 4<br />

STD 5 CONC. RACK POS. 558 - 5<br />

STD 6 CONC. RACK POS. 0.0 - 6<br />

UNIT<br />

ng/ml.<br />

SD LIMIT 300<br />

DUPLICATE LIMIT 500<br />

SENSITIVITY 0<br />

ABS. LIMIT<br />

0 - INCREASE<br />

PROZONE LIMIT<br />

0 - LOWER<br />

EXPECTED VALUE 15 - 300<br />

INSTRUMENT FACTOR 1.0<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

CRP Turbidymetric<br />

Preparation of working reagent:<br />

Follow instructions given on the procedure.<br />

Calibration:<br />

Use CRP CALIBRATOR.<br />

Operation conditions:<br />

Non-Linear Calibration.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

CRP<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE [2 POINTS] - 20 - 32<br />

SAMPLE VOLUME (µl) 6<br />

R1 VOLUME (µl) 350<br />

R2 VOLUME (µl) 50<br />

WAVELENGHT 505<br />

CALIB. METHOD NON-LINEAR 1 - 5<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. 5 - 2<br />

STD 3 CONC. RACK POS. 10 - 3<br />

STD 4 CONC. RACK POS. 20 - 4<br />

STD 5 CONC. RACK POS. 40 - 5<br />

STD 6 CONC. RACK POS. 0.0 - 6<br />

UNIT<br />

mg/l.<br />

SD LIMIT 300<br />

DUPLICATE LIMIT 1000<br />

SENSITIVITY 0<br />

ABS. LIMIT<br />

0 - INCREASE<br />

PROZONE LIMIT<br />

0 - LOWER<br />

EXPECTED VALUE 0 - 8<br />

INSTRUMENT FACTOR 1.0<br />

#Enhanced sensitivity (i.e. paediatric samples).<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

CRP Turbidymetric<br />

Preparation of working reagent:<br />

Follow instructions given on the procedure.<br />

Calibration:<br />

Use CRP CALIBRATOR.<br />

Operation conditions:<br />

Linear Calibration.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

CRP<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE [2 POINTS] - 20 - 32<br />

SAMPLE VOLUME (µl) 3<br />

R1 VOLUME (µl)<br />

350 - 20 - NO<br />

R2 VOLUME (µl)<br />

50 - 20 - NO<br />

WAVELENGHT 505<br />

CALIB. METHOD NON-LINEAR 1 - 5<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. 10 - 2<br />

STD 3 CONC. RACK POS. 20 - 3<br />

STD 4 CONC. RACK POS. 40 - 4<br />

STD 5 CONC. RACK POS. 80 - 5<br />

STD 6 CONC. RACK POS. 0.0 - 6<br />

UNIT<br />

mg/l.<br />

SD LIMIT 300<br />

DUPLICATE LIMIT 1000<br />

SENSITIVITY 0<br />

ABS. LIMIT<br />

0 - INCREASE<br />

PROZONE LIMIT<br />

0 - LOWER<br />

EXPECTED VALUE 0 - 8<br />

INSTRUMENT FACTOR 1.0<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

CREATININE<br />

Preparation of working reagent:<br />

The reagents are ready to use. Alkaline reagent is used as reagent 1 and picric reagent is used as reagent<br />

2.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

CREA<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 5 - 19 - 21<br />

SAMPLE VOLUME (µl) 20<br />

R1 VOLUME (µl) 175 - 50 - 0<br />

R2 VOLUME (µl) 175 - 50 - 0<br />

WAVELENGHT 600 - 505<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

mg/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 10000 - 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 0.7 - 1.4<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 15 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

CHOLINESTERASE<br />

Preparation of working reagent:<br />

Dissolve one tablet into one vial of buffer. The working reagent is stable 2 hours refrigerated.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

COLIN<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 5 - 5 - 9<br />

SAMPLE VOLUME (µl) 2<br />

R1 VOLUME (µl) 500 - 20 - 0<br />

R2 VOLUME (µl) 0 - 50 - 0<br />

WAVELENGHT 660 - 415<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

U/L.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 20000 - 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 4659 - 14443<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 15000 U/L.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

CHOLESTEROL<br />

Preparation of working reagent:<br />

Dissolve one vial R2 with a little quantity of buffer. Return to the original vial when it is completely<br />

dissolved. The working reagent is stable 4 months refrigerated and out of the light.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

CHOL<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 31 - 0<br />

SAMPLE VOLUME (µl) 4<br />

R1 VOLUME (µl) 400 - 50 - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 600 - 505<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

mg/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 120 - 260<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 600 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

CHLORIDE<br />

Preparation of working reagent:<br />

The reagent is ready to use.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

CLOR<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 31 - 0<br />

SAMPLE VOLUME (µl) 5<br />

R1 VOLUME (µl) 400 - 50 - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 660 - 505<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

mEQ/L.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 98 - 107<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 130 mEQ/L.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

CALCIUM OCC<br />

Preparation of working reagent:<br />

The reagents are ready to use. Buffer is used as R1 and colour solution as R2.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

CALC<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 2 - 15 - 32<br />

SAMPLE VOLUME (µl) 5<br />

R1 VOLUME (µl) 250 - [...] - 0<br />

R2 VOLUME (µl) 250 - [...] - 0<br />

WAVELENGHT 660 - 570<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

mg/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 9.0 - 11.0<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 15 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

β2-Microglobulin Turbidymetric<br />

Preparation of working reagent:<br />

Follow instructions given on the procedure.<br />

Calibration:<br />

Use Beta-2-Microg. CALIBRATOR.<br />

Operation conditions:<br />

Non-Linear Calibration.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

BETA2<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE [2 POINTS] - 20 - 32<br />

SAMPLE VOLUME (µl) 3<br />

R1 VOLUME (µl)<br />

350 - 20 - NO<br />

R2 VOLUME (µl)<br />

50 - 20 - NO<br />

WAVELENGHT 600<br />

CALIB. METHOD NON-LINEAR 1 - 5<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. 0.68 - 2<br />

STD 3 CONC. RACK POS. 1.35 - 3<br />

STD 4 CONC. RACK POS. 2.70 - 4<br />

STD 5 CONC. RACK POS. 5.40 - 5<br />

STD 6 CONC. RACK POS. 10.8 - 6<br />

UNIT<br />

mg/l.<br />

SD LIMIT 300<br />

DUPLICATE LIMIT 1000<br />

SENSITIVITY 0<br />

ABS. LIMIT<br />

0 - INCREASE<br />

PROZONE LIMIT<br />

0 - LOWER<br />

EXPECTED VALUE 0.8 - 2.4<br />

INSTRUMENT FACTOR 1.0<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

ASO Turbidymetric<br />

Preparation of working reagent:<br />

Follow instructions given on the procedure.<br />

Calibration:<br />

Use ASO CALIBRATOR.<br />

Operation conditions:<br />

Non-Linear Calibration.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

ASO<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE [2 POINTS] - 20 - 32<br />

SAMPLE VOLUME (µl) 3<br />

R1 VOLUME (µl)<br />

350 - 20 - NO<br />

R2 VOLUME (µl)<br />

40 - 20 - NO<br />

WAVELENGHT 505<br />

CALIB. METHOD NON-LINEAR 1 - 5<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. 100 - 2*<br />

STD 3 CONC. RACK POS. 200 – 3<br />

STD 4 CONC. RACK POS. 400 – 4<br />

STD 5 CONC. RACK POS. 800 – 5<br />

STD 6 CONC. RACK POS. 0.0 – 6<br />

UNIT<br />

IU/ml.<br />

SD LIMIT 300<br />

DUPLICATE LIMIT 1000<br />

SENSITIVITY 0<br />

ABS. LIMIT<br />

0 - INCREASE<br />

PROZONE LIMIT<br />

0 - LOWER<br />

EXPECTED VALUE 0 - 250<br />

INSTRUMENT FACTOR 1.0<br />

(*) Dilute standard ASO 1+1 with 0.9% NaCl saline solution to give a standard concentration of 400 IU/ml.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

ASO Turbidymetric<br />

Preparation of working reagent:<br />

Follow instructions given on the procedure.<br />

Calibration:<br />

Use ASO CALIBRATOR.<br />

Operation conditions:<br />

Linear Calibration.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

ASO<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE [2 POINTS] - 20 - 32<br />

SAMPLE VOLUME (µl) 3<br />

R1 VOLUME (µl)<br />

350 - 20 - NO<br />

R2 VOLUME (µl)<br />

40 - 20 - NO<br />

WAVELENGHT 505<br />

CALIB. METHOD LINEAR 0 - 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. 400 - 2*<br />

STD 3 CONC. RACK POS. 0.0 – 3<br />

STD 4 CONC. RACK POS. 0.0 – 4<br />

STD 5 CONC. RACK POS. 0.0 – 5<br />

STD 6 CONC. RACK POS. 0.0 – 6<br />

UNIT<br />

IU/ml.<br />

SD LIMIT 300<br />

DUPLICATE LIMIT 1000<br />

SENSITIVITY 0<br />

ABS. LIMIT<br />

0 - INCREASE<br />

PROZONE LIMIT<br />

0 - LOWER<br />

EXPECTED VALUE 0 - 250<br />

INSTRUMENT FACTOR 1.0<br />

(*) Dilute standard ASO 1+1 with 0.9% NaCl saline solution to give a standard concentration of 400 IU/ml.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

ALKALINE PHOSPHATASE<br />

Preparation of working reagent:<br />

Dissolve one tablet into one vial of buffer or one tablet with 15 ml. of buffer, depending of presentation.<br />

The working reagent is stable 2 days refrigerated.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

AKP<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 5 - 15 - 25<br />

SAMPLE VOLUME (µl) 20<br />

R1 VOLUME (µl) 300 - 20 - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 660 - 415<br />

CALIB. METHOD 2 - 0<br />

STD 1 CONC. RACK POS. 0 - 1<br />

UNIT<br />

U/L<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 15000 - 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 0.0 - 5.4<br />

INSTRUMENT FACTOR 1.00<br />

3-CALIBRATION MONITOR<br />

K=<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 150 U/L.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

αAMYLASE<br />

Preparation of working reagent:<br />

Dissolve one tablet into one vial of buffer or one tablet with 15 ml. of buffer, depending of presentation.<br />

The working reagent is stable 21 days refrigerated.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

AMIL<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 5 - 5 - 10<br />

SAMPLE VOLUME (µl) 10<br />

R1 VOLUME (µl) 350 - [....] - 0<br />

R2 VOLUME (µl) 0 - 20 - 0<br />

WAVELENGHT 660 - 415<br />

CALIB. METHOD 2 - 0<br />

STD 1 CONC. RACK POS. 0 – 1<br />

UNIT<br />

U/L.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 15000 - 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 0 - 90<br />

INSTRUMENT FACTOR 1.00<br />

3-CALIBRATION MONITOR<br />

K=<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 1500 U/L.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

URIC ACID<br />

Preparation of working reagent:<br />

Dissolve one vial of lyophilised R2 with a little quantity of buffer R1. When it's dissolved completely retune<br />

to the original flack of buffer.<br />

Calibration:<br />

Use LABTROL serum calibrator.<br />

Operation conditions:<br />

Push on the key PARAMETER JOB and introduce the following information into the programs 1, 2, and 4,<br />

and then continue following instructions of instrument.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

URIC<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE 1 - 31 - 0<br />

SAMPLE VOLUME (µ l) 10<br />

R1 VOLUME (µl) 400 - [....] - 0<br />

R2 VOLUME (µl) 0 – 1 – 0<br />

WAVELENGHT 600 – 505<br />

CALIB. METHOD 1 – 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. [...] - 2<br />

UNIT<br />

mg/dl.<br />

SD LIMIT 0.1<br />

DUPLICATE LIMIT 100<br />

SENSITIVITY 0<br />

ABS. LIMIT 0 – 0<br />

PROZONE LIMIT 0<br />

EXPECTED VALUE 24 – 7.0<br />

INSTRUMENT FACTOR 1.00<br />

4- CHANEL ASSIGNEMENT<br />

PHASE [....]<br />

TEST [....]<br />

Linearity:<br />

The method is linear until 25 mg/dl.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

CRP Turbi.<br />

Preparation of working reagent:<br />

Follow instructions given on the procedure.<br />

Calibration:<br />

Use CRP CALIBRATOR.<br />

Operation conditions:<br />

Linear Calibration.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

CRP<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE [2 POINTS] - 20 - 32<br />

SAMPLE VOLUME (µl) 3<br />

R1 VOLUME (µl)<br />

350 - 20 - NO<br />

R2 VOLUME (µl)<br />

50 - 20 - NO<br />

WAVELENGHT 505<br />

CALIB. METHOD NON-LINEAR 1 - 5<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. 10 - 2<br />

STD 3 CONC. RACK POS. 20 - 3<br />

STD 4 CONC. RACK POS. 40 - 4<br />

STD 5 CONC. RACK POS. 80 - 5<br />

STD 6 CONC. RACK POS. 0.0 - 6<br />

UNIT<br />

mg/l.<br />

SD LIMIT 300<br />

DUPLICATE LIMIT 1000<br />

SENSITIVITY 0<br />

ABS. LIMIT<br />

0 - INCREASE<br />

PROZONE LIMIT<br />

0 - LOWER<br />

EXPECTED VALUE 0 - 8<br />

INSTRUMENT FACTOR 1.0<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

ASO Turbi.<br />

Preparation of working reagent:<br />

Follow instructions given on the procedure.<br />

Calibration:<br />

Use ASO CALIBRATOR.<br />

Operation conditions:<br />

Non-Linear Calibration.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

ASO<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE [2 POINTS] - 20 - 32<br />

SAMPLE VOLUME (µl) 3<br />

R1 VOLUME (µl)<br />

350 - 20 - NO<br />

R2 VOLUME (µl)<br />

40 - 20 - NO<br />

WAVELENGHT 505<br />

CALIB. METHOD NON-LINEAR 1 - 5<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. 100 - 2*<br />

STD 3 CONC. RACK POS. 200 – 3<br />

STD 4 CONC. RACK POS. 400 – 4<br />

STD 5 CONC. RACK POS. 800 – 5<br />

STD 6 CONC. RACK POS. 0.0 – 6<br />

UNIT<br />

IU/ml.<br />

SD LIMIT 300<br />

DUPLICATE LIMIT 1000<br />

SENSITIVITY 0<br />

ABS. LIMIT<br />

0 - INCREASE<br />

PROZONE LIMIT<br />

0 - LOWER<br />

EXPECTED VALUE 0 - 250<br />

INSTRUMENT FACTOR 1.0<br />

(*) Dilute standard ASO 1+1 with 0.9% NaCl saline solution to give a standard concentration of 400 IU/ml.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

ASO Turbi.<br />

Preparation of working reagent:<br />

Follow instructions given on the procedure.<br />

Calibration:<br />

Use ASO CALIBRATOR.<br />

Operation conditions:<br />

Linear Calibration.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

ASO<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE [2 POINTS] - 20 - 32<br />

SAMPLE VOLUME (µl) 3<br />

R1 VOLUME (µl)<br />

350 - 20 - NO<br />

R2 VOLUME (µl)<br />

40 - 20 - NO<br />

WAVELENGHT 505<br />

CALIB. METHOD LINEAR 0 - 0<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. 400 - 2*<br />

STD 3 CONC. RACK POS. 0.0 – 3<br />

STD 4 CONC. RACK POS. 0.0 – 4<br />

STD 5 CONC. RACK POS. 0.0 – 5<br />

STD 6 CONC. RACK POS. 0.0 – 6<br />

UNIT<br />

IU/ml.<br />

SD LIMIT 300<br />

DUPLICATE LIMIT 1000<br />

SENSITIVITY 0<br />

ABS. LIMIT<br />

0 - INCREASE<br />

PROZONE LIMIT<br />

0 - LOWER<br />

EXPECTED VALUE 0 - 250<br />

INSTRUMENT FACTOR 1.0<br />

(*) Dilute standard ASO 1+1 with 0.9% NaCl saline solution to give a standard concentration of 400 IU/ml.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

CRP Turbi.<br />

Preparation of working reagent:<br />

Follow instructions given on the procedure.<br />

Calibration:<br />

Use CRP CALIBRATOR.<br />

Operation conditions:<br />

Non-Linear Calibration.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

CRP<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE [2 POINTS] - 20 - 32<br />

SAMPLE VOLUME (µl) 6<br />

R1 VOLUME (µl) 350<br />

R2 VOLUME (µl) 50<br />

WAVELENGHT 505<br />

CALIB. METHOD NON-LINEAR 1 - 5<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. 5 - 2<br />

STD 3 CONC. RACK POS. 10 - 3<br />

STD 4 CONC. RACK POS. 20 - 4<br />

STD 5 CONC. RACK POS. 40 - 5<br />

STD 6 CONC. RACK POS. 0.0 - 6<br />

UNIT<br />

mg/l.<br />

SD LIMIT 300<br />

DUPLICATE LIMIT 1000<br />

SENSITIVITY 0<br />

ABS. LIMIT<br />

0 - INCREASE<br />

PROZONE LIMIT<br />

0 - LOWER<br />

EXPECTED VALUE 0 - 8<br />

INSTRUMENT FACTOR 1.0<br />

#Enhanced sensitivity (i.e. paediatric samples).<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002


<strong>LABKIT</strong> HITACHI 704<br />

RF Turbi.<br />

Preparation of working reagent:<br />

Follow instructions given on the procedure.<br />

Calibration:<br />

Use RF CALIBRATOR.<br />

Operation conditions:<br />

Non-Linear Calibration.<br />

1- TEST NAME<br />

TEST [.....]<br />

TEST NAME<br />

RF<br />

2- CHEMISTRY PARAMETERS<br />

TEST [.....]<br />

ASSAY CODE [2 POINTS] - 20 - 32<br />

SAMPLE VOLUME (µl) 3<br />

R1 VOLUME (µl)<br />

350 - 20 - NO<br />

R2 VOLUME (µl)<br />

40 - 20 - NO<br />

WAVELENGHT 600<br />

CALIB. METHOD NON-LINEAR 1 - 5<br />

STD 1 CONC. RACK POS. 0.0 – 1<br />

STD 2 CONC. RACK POS. 16.5 – 2*<br />

STD 3 CONC. RACK POS. 33 – 3*<br />

STD 4 CONC. RACK POS. 66 – 4*<br />

STD 5 CONC. RACK POS. 132 – 5*<br />

STD 6 CONC. RACK POS. 0.0 – 6<br />

UNIT<br />

IU/ml.<br />

SD LIMIT 300<br />

DUPLICATE LIMIT 1000<br />

SENSITIVITY 0<br />

ABS. LIMIT<br />

0 - INCREASE<br />

PROZONE LIMIT<br />

0 - LOWER<br />

EXPECTED VALUE 0 - 20<br />

INSTRUMENT FACTOR 1.0<br />

*Depends on the concentration of calibrated used.<br />

CHEMELEX, S.A. Polígono Industrial Can Castells. C/ Industria, 113 nave J 08420 Canovelles Tel/ +34 93 849 17 35<br />

Fax/ +34 93 846 78 75 e-mail: chemelex@terra.es<br />

Ed. 10.2002