SERIES <strong>3000</strong> MKII TECHNICAL MANUAL<br />
S3KMKIIMAN_MAN0878_Issue 1_01-10<br />
4. Put the magnet over the ‘’ switch and remove to enter the Calibration menu.<br />
5. The display will show the current gas reading, ‘ZEro’ and the ‘ ’ icon flashes.<br />
6. When the zero gas reading is stable use ‘’ to confirm zero calibration.<br />
7. If successful the display shows ‘ZEro PASS’ (if not successful, the display<br />
shows ‘ZEro FAIL’ and returns to menu mode).<br />
8. If using zero-air, turn it off. Zeroing is complete and saved.<br />
9. The display shows ‘SPAn’ with ‘YES’ flashing.<br />
10. If span calibration is required use ‘’ proceed to the next step. If span<br />
calibration is not required, use ‘’ to select ‘No’ and ‘’ to return to menu<br />
mode.<br />
11. The display shows the current calibration span gas concentration flashing. Use<br />
‘’ to change the calibration span gas concentration, and ‘’ when required<br />
span calibration level is set.<br />
12. Span calibration point is displayed with ‘YES’ flashing. Use ‘’ to confirm or<br />
‘’ to select ‘No’ and return and enter a new span calibration point.<br />
13. The display will show the current gas reading, ‘SPAN’ and the ‘ ’ icon flashes.<br />
14. Connect the regulator to the span gas cylinder.<br />
15. Apply the span gas to the sensor using the calibration gas flow housing. The<br />
live gas reading is displayed. When the reading is stable, use ‘’ to confirm<br />
span calibration.<br />
16. If the sensor cell has been replaced the following display may be shown.<br />