Doppler Insonification Angle = 30°<br />

The angle formed between the flow direction and the steered Doppler line<br />

is 30 degrees. Notice how the flow is traveling closer to the transducer,<br />

but not directly toward as when the angle was 0 degrees. As a result, less<br />

of the Doppler shift is detected at 30 degrees than at 0 degrees, but the<br />

shift is still positive.