07.01.2015 Views

CR200/CR200X Series Dataloggers - Campbell Scientific

CR200/CR200X Series Dataloggers - Campbell Scientific

CR200/CR200X Series Dataloggers - Campbell Scientific

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Section 4. Sensor Support<br />

4.3.1 Measurements Requiring AC Excitation<br />

Some resistive sensors require AC Excitation. These include electrolytic tilt<br />

sensors, soil moisture blocks, water conductivity sensors, and wetness sensing<br />

grids. The use of DC excitation in these sensors can result in polarization, which<br />

will cause erroneous measurement, shift calibration, or lead to rapid sensor<br />

decay.<br />

Other sensors, e.g., LVDTs (Linear Variable Differential Transformer), require<br />

and AC excitation because they rely on inductive coupling to provide a signal.<br />

DC excitation will provide no output.<br />

<strong>CR200</strong>(X) bridge measurements cannot reverse excitation polarity to provide<br />

AC excitation and avoid ion polarization. Sensors requiring AC excitation<br />

should not be used with the <strong>CR200</strong>(X).<br />

Other <strong>Campbell</strong> <strong>Scientific</strong> dataloggers (e.g. CR800 series, CR1000, CR3000)<br />

are compatible with sensors that require AC excitation.<br />

4.4 Pulse Count Measurement<br />

FIGURE. Switch Closure Pulse Sensor p. 42 is a generalized schematic showing<br />

connection of a pulse sensor to the <strong>CR200</strong>(X). The <strong>CR200</strong>(X) features two<br />

dedicated pulse input channels, P_SW and P_LL, and two digital I/O channels,<br />

C1and C2, for measuring pulse output sensors. Activated by the PulseCount ()<br />

instruction, dedicated 16-bit counters on P_SW, P_LL, C1 and C2 are used to<br />

accumulate all counts over the user specified scan interval. The value which is<br />

output for each scan is the difference in the last known counter value and the<br />

new counter value. Since the last count is maintained for each scan, even if the<br />

counter rolls over between scans the correct count will be recorded. If the time<br />

between scans is such that the counter exceeds 65,536 pulses during a scan, then<br />

the counter will roll over twice resulting is an erroneous measurement.<br />

PulseCount () instruction parameters specify the pulse input type, channel used,<br />

and pulse output option.<br />

Note: The PulseCount instruction must be executed once before the pulse or<br />

control port is ready for input. This may be of particular concern for programs<br />

with long scan intervals. For example, the PulseCount () instruction will not<br />

yield a valid output until the turn of the second hour if the PulseCount ()<br />

instruction is used within a program with a scan interval of 1 hour.<br />

Figure 29: Switch Closure Pulse Sensor<br />

42

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!