Section 4. Sensor Support<br />

Figure 27: Voltage Measurement Accuracy (0° to 40° C)<br />

4.2.3 Voltage Range<br />

The <strong>CR200</strong>(X) has one analog voltage range of 0 to 2.5 volts. The resolution<br />

for a single A/D conversion is 0.6 millivolts.<br />

4.2.4 Integration<br />

Integration is used to reduce the noise included in a measurement. The<br />

<strong>CR200</strong>(X) uses a form of digital integration. It makes 10 A/D conversions and<br />

averages them for the result returned. The A/D conversions are made every 26<br />

microseconds.<br />

Averaging will also reduce signal noise from the sensor such as when water<br />

level measured with a pressure transducer is changing because of pressure<br />

fluctuations caused by wind.<br />

Averaging also has an effect on the resolution. The resolution seen in the<br />

numerical result is the resolution of a single A/D conversion (0.6 mV) divided<br />

by the number of A/D conversions (10).<br />

40