Instruction Manual<br />

IBRit-md2/md2s/md2c<br />

5.1 Self Check Routine<br />

The IBRit-md2/-md2s/-md2c self check routine is invoked at power-on and on receipt of a RESET<br />

command. All system components are checked. The test result is signalled by the IT-lamp on the rear<br />

panel. Four short flashes indicate no errors were found. A sequences of long and short flashes indicate<br />

errors. The table gives the meanings of the error signals. The IBRit-md2/-md2s/-md2c remains in the self<br />

test mode until the fault is cleared and cannot be used for gauging. The self-check cycle takes 3 to 4<br />

seconds.<br />

Self-check Error Flash Sequence :<br />

-------------------------------------------------------------------------------------------------------------------------------------<br />

Test result : Flash sequence : Error :<br />

-------------------------------------------------------------------------------------------------------------------------------------<br />

--- S S S S No errors detected, self-check complete<br />

Error 1 L S S S Error in programme memory<br />

Error 2 S L S S Error in working memory<br />

Notice : S - short flash ( 0,5 seconds )<br />

L - long flash ( 1 second )<br />