Fill up water tank<br />

Switch on H-<strong>Cube®</strong><br />

Insert CatCart®<br />

CONTROLLED MODE<br />

Wait until hydrogen is<br />

eluting from the reaction<br />

line and the hydrogen<br />

output is constantly under<br />

25 in the bubble detec<strong>to</strong>r<br />

box on the Service Screen<br />

Flow Solvent<br />

Through System<br />

Wait until you see solvent eluting<br />

from the reaction line!!!<br />

Set Temperature and Pressure and Mode<br />

on <strong>to</strong>uch Screen, Press START<br />

Transfer pump inlet line over<br />

<strong>to</strong> the reaction mixture<br />

Wait until hydrogen comes<br />

out of the reaction line<br />

• For reactions at 1 bar-Use Full Hydrogen Mode<br />

• For best results in controlled mode start at 50 bar<br />

• For solvent compatability-see page 66 of the manual<br />

• For information on system blockages-see p. 55 of the manual<br />

FULL HYDROGEN MODE<br />

When the reaction mixture is<br />

almost dry, add reaction<br />

solvent-Wash for 10 mins.<br />

When you have finished<br />

flushing the system with<br />

solvent press S<strong>to</strong>p or S<strong>to</strong>p<br />

and Keep H2 on the <strong>to</strong>uchscreen<br />

and S<strong>to</strong>p on the pump