Section 7. Installation<br />

lightning strike. Figure Lightning-Protection Scheme (p. 89) shows a simple<br />

lightning-protection scheme utilizing a lightning rod, metal mast, heavy-gage<br />

ground wire, and ground rod to direct damaging current away from the <strong>CR1000</strong>.<br />

Figure 32: Lightning-protection scheme<br />

7.5.2 Single-Ended Measurement Reference<br />

Low-level, single-ended voltage measurements are sensitive to ground potential<br />

fluctuations. The grounding scheme in the <strong>CR1000</strong> has been designed to eliminate<br />

ground potential fluctuations due to changing return currents from 12V, SW-12,<br />

5V, and C1 – C8 terminals. This is accomplished by utilizing separate signal<br />

89