Section 10. Troubleshooting<br />

Figure 135: DevConfig terminal emulator tab<br />

10.6.1 Serial Talk Through and Sniffer<br />

In the P: Serial Talk Through and W: Serial Comms Sniffer modes, the<br />

timeout can be changed from the default of 40 seconds to any value ranging from<br />

1 to 86400 seconds (86400 seconds = 1 day).<br />

When using options P or W in a terminal session, consider the following:<br />

1. Concurrent terminal sessions are not allowed by the <strong>CR1000</strong>.<br />

2. Opening a new terminal session will close the current terminal session.<br />

3. The <strong>CR1000</strong> will attempt to enter a terminal session when it receives non-<br />

PakBus characters on the nine-pin RS-232 port or CS I/O port, unless the port<br />

is first opened with the SerialOpen() command.<br />

If the <strong>CR1000</strong> attempts to enter a terminal session on the nine-pin RS-232 port or<br />

CS I/O port because of an incoming non-PakBus character, and that port was not<br />

opened using the SerialOpen() command, any currently running terminal<br />

function, including the communication sniffer, will immediately stop. So, in<br />

programs that frequently open and close a serial port, the probability is higher that<br />

a non-PakBus character will arrive at the closed serial port, thus closing an<br />

existing talk-through or sniffer session.<br />