Section 10. Troubleshooting<br />

Charging Regulator with Solar-Panel Test<br />

Disconnect any wires attached to the 12V and G (ground) terminals on the PS100 or CH100 charging regulator. Unplug any batteries. Connect the solar panel to the two<br />

CHG terminals. Polarity of inputs does not matter. Only the solar panel should be connected. Set the charging-regulator power switch to OFF.<br />

NOTE This test assumes the solar panel has an unregulated output.<br />

Solar Panel Test<br />

Set a voltmeter to measure DC voltage. Measure solar panel output<br />

across the two solar-panel leads by placing a voltmeter lead on one<br />

CHG terminal, and the other lead on the other CHG terminal. Is the<br />

output 17 to 22 Vdc<br />

Yes<br />

No<br />

Remove the solar-panel leads from the<br />

charging circuit. Measure solar-panel<br />

output across the two leads. Is the<br />

output > 0 Vdc<br />

No<br />

The solar panel is damaged<br />

and should be repaired or<br />

replaced.<br />

Yes<br />

Is the voltage ≥ 17 Vdc<br />

No<br />

There may not be enough<br />

sunlight to perform the test,<br />

or the solar panel is<br />

damaged.<br />

Yes<br />

Yes<br />

Reconnect the power source<br />

(transformer / solar panel) to the CHG<br />

terminals on the charging regulator.<br />

Measure the voltage between the two<br />

CHG terminals. Is the voltage ≥ 17 Vdc<br />

/ Vac<br />

No<br />

5-kΩ Load Test<br />

1) Place a 5-kΩ resistor between a 12V terminal and a G (ground)<br />

terminal on the charging regulator.<br />

2) Switch the power switch to ON.<br />

3) Measure the DC voltage across the resistor.<br />

Is the measured voltage 13.3 to 14.1 V<br />

No<br />

Measure the voltage between the two<br />

pins in a battery-connection receptacle.<br />

Is the voltage 10.0 to 15.5 Vdc<br />

No<br />

Yes<br />

Yes<br />

50-Ω Load Test<br />

1) Switch the power switch to OFF.<br />

2) Disconnect the power source (transformer / solar panel).<br />

3) Remove the 5-kΩ resistor<br />

4) Place a 50-Ω, 1-Watt resistor between a 12V terminal and a G<br />

(ground) terminal on the charging regulator.<br />

5) Reconnect the power source and then switch the power switch to<br />

ON.<br />

7) Measure the voltage across the ends of the resistor.<br />

Is the voltage 13.0 to 14.0 Vdc (13.3 if circuit just adjusted)<br />

8) Switch the power switch to OFF.<br />

NOTE The resistor will get HOT in just a few seconds. After<br />

measuring the voltage, switch the power switch to OFF and allow<br />

the resistor to cool before removing it.<br />

Yes<br />

No<br />

See Adjusting Charging Voltage (p.<br />

440) to calibrate the charging regulator,<br />

or return the charging regulator to<br />

<strong>Campbell</strong> <strong>Scientific</strong> for calibration.<br />

With the charging regulator still under<br />

load, measure the voltage between the<br />

two CHG terminals. Is the voltage ><br />

15.5 Vdc<br />

Yes<br />

Obtain Repair<br />

Authorization<br />

The charging regulator is<br />

damaged and should be<br />

repaired or replaced.<br />

No<br />

Test Completed<br />

The charger is functioning<br />

properly. Remove the 50-Ω<br />

resistor.<br />

There may not be enough sunlight to<br />

perform the test.<br />