Section 8. Operation<br />

Figure 102: Input conditioning circuit for period averaging<br />

8.1.7 SDI-12 Recording<br />

8.1.8 RS-232 and TTL<br />

Read More! SDI-12 Sensor Support (p. 172) and Serial Input / Output (p. 509).<br />

SDI-12 is a communications protocol developed to transmit digital data from<br />

smart sensors to data-acquisition units. It is a simple protocol, requiring only a<br />

single communication wire. Typically, the data-acquisition unit also supplies<br />

power (12 Vdc and ground) to the SDI-12 sensor. The <strong>CR1000</strong> is equipped with 4<br />

SDI-12 channels (C1, C3, C5, C7) and an SDI12Recorder() CRBasic instruction.<br />

Read More! Serial Input / Output Instructions (p. 509) and Serial I/O (p. 200).<br />

The <strong>CR1000</strong> can usually receive and record RS-232 and 0 – 5 Vdc logic data<br />

from sensors designed to transmit via these protocols. Data are received through<br />

the CS I/O port with the proper interface (see the appendix CS I/O Serial<br />

Interfaces (p. 567) ), the RS-232 port, or the digital I/O communication ports (C1 &<br />

C2, C3 & C4, C5 & C6, C7 & C8). If additional serial inputs are required, serial<br />

input expansion modules (see the appendix Serial Input Expansion Modules ) can<br />

be connected to increase the number of serial ports. Serial data are usually<br />

captured as text strings, which are then parsed (split up) as defined in the user<br />

entered program.<br />

Note Digital I/O communication ports (control ports) only transmit 0 – 5 Vdc<br />

logic. However, they read most true RS-232 input signals. When connecting serial<br />

sensors to an Rx control port, the sensor power consumption may increase by a<br />

few milliamps due to voltage clamps. An external resistor may need to be added<br />

in series to the Rx line to limit the current drain, although this is not advisable at<br />

very high baud rates. Figure Circuit to Limit Control Port Input to 5 Volts (p.<br />

324) shows a circuit that limits voltage input on a control port to 5 Vdc.<br />

323