Section 8. Operation<br />

Table 73. Time Constants (τ)<br />

Measurement<br />

τ<br />

Pulse channel, high-frequency mode 1.2<br />

Pulse channel, switch-closure mode 3300<br />

Pulse channel, low-level ac mode<br />

See table Filter Attenuation of<br />

Frequency Signals (p. 321) footnote<br />

Digital I/O, high-frequency mode 0.025<br />

Digital I/O, switch-closure mode 0.025<br />

Table 74. Filter Attenuation of Frequency Signals.<br />

As shown for low-level ac inputs, increasing voltage is required at increasing frequencies to<br />

overcome filter attenuation on pulse-input channels*.<br />

ac mV (RMS)<br />

20<br />

200<br />

2000<br />

5000<br />

Maximum Frequency<br />

20<br />

200<br />

10,000<br />

20,000<br />

*8.5-ms time constant filter (19 Hz 3 dB frequency) for low-amplitude signals. 1-ms time<br />

constant (159 Hz 3 dB frequency) for larger (> 0.7 V) amplitude signals.<br />

Figure 101: Amplitude reduction of pulse-count waveform (before and after<br />

1-µs time constant filter)<br />

321