04.01.2015 Views

CR1000 Manual - Campbell Scientific

CR1000 Manual - Campbell Scientific

CR1000 Manual - Campbell Scientific

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Section 8. Operation<br />

8.1.5.4 Pulse Measurement Problems<br />

8.1.5.4.1 Pay Attention to Specifications<br />

The table Example of Differing Specifications for Pulse Input Channels (p. 320)<br />

compares specifications for pulse-input channels to emphasize the need for<br />

matching the proper device to application. Take time to understand signals to be<br />

measured and compatible channels.<br />

Table 72. Example of Differing Specifications for Pulse-Input<br />

Channels<br />

Pulse Channels<br />

P1, P2<br />

Digital I/O Channels<br />

C1, C2, C3, C4,<br />

C5, C6, C7, C8<br />

High Frequency Max 250 kHz 400 kHz<br />

Max Input Voltage 20 Vdc 16 Vdc<br />

State Transition<br />

Thresholds<br />

Count upon transition from<br />

2.2 Vdc<br />

Count upon transition from<br />

3.8 Vdc<br />

8.1.5.4.2 Input Filters and Signal Attenuation<br />

Pulse-input channels are equipped with input filters to reduce spurious noise that<br />

can cause false counts. The higher the time constant (τ) of the filter, the tighter<br />

the filter. Table Time Constants (p. 321) lists τ values for pulse-input channels. So,<br />

while TimerIO() frequency measurement may be superior for clean signals, a<br />

pulse channel filter (much higher τ) may be required to get a measurement on a<br />

dirty signal.<br />

Input filters, however, attenuate the amplitude (voltage) of the signal. The<br />

amount of attenuation is a function of the frequency passing through the filter.<br />

Higher-frequency signals are attenuated more. If a signal is attenuated enough, it<br />

may not pass the state transition thresholds required by the detection device (listed<br />

in table Pulse-Input Channels and Measurements (p. 39) ). To avoid over<br />

attenuation, sensor output voltage must be increased at higher frequencies. As an<br />

example, table Filter Attenuation of Frequency Signals (p. 321) lists low-level ac<br />

frequencies and the voltages required to overcome filter attenuation.<br />

For pulse-input channels P1 – P2, an RC input filter with an approximate 1-μs<br />

time constant precedes the inverting CMOS input buffer. The resulting amplitude<br />

reduction is illustrated in figure Amplitude Reduction of Pulse-Count Waveform (p.<br />

321). For a 0- to 5-Vdc square wave applied to a pulse channel, the maximum<br />

frequency that can be counted in high-frequency mode is approximately 250 kHz.<br />

320

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!