Section 8. Operation<br />

outside the chamber. The temperature of this bar was also measured by another<br />

datalogger. Differences between the temperature measured by one of the<br />

thermocouples and the actual temperature of the bar are due to the temperature<br />

difference between the terminals the thermocouple is connected to and the<br />

thermistor reference (the figures have been corrected for thermistor errors). Figure<br />

Panel-Temperature Gradients (Low Temperature to High) (p. 304) shows the errors<br />

when the chamber was changed from low temperature to high in approximately 15<br />

minutes. Figure Panel-Temperature Gradients (High Temperature to Low) (p. 304)<br />

shows the results when going from high temperature to low. During rapid<br />

temperature changes, the panel thermistor will tend to lag behind terminal<br />

temperature because it is mounted deeper in the <strong>CR1000</strong>.<br />

Figure 92: Panel-temperature error summary<br />

303