Section 8. Operation<br />

Table 57. CRBasic Measurement Settling Times<br />

Settling<br />

Time<br />

Entry<br />

Input<br />

Voltage<br />

Range<br />

Integration<br />

Code<br />

Settling<br />

Time 1<br />

0 All 250 450 µs (default)<br />

0 All _50Hz 3 ms (default)<br />

0 All _60Hz 3 ms (default)<br />

>100 All X 2 μs entered<br />

1 Minimum settling time required to allow the input to settle to <strong>CR1000</strong> resolution specifications.<br />

2 X is an integer >100.<br />

8.1.2.8.1 Minimizing Settling Errors<br />

A settling time is required for voltage measurements to minimize the effects of the<br />

following sources of error:<br />

• A small switching transient occurs when the <strong>CR1000</strong> switches to the singleended<br />

or differential channel to be measured.<br />

• A relatively large transient may be induced on the signal conductor via<br />

capacitive coupling during a bridge measurement from an adjacent excitation<br />

conductor.<br />

• 50-Hz or 60-Hz integrations require a relatively long reset time of the internal<br />

integration capacitor before the next measurement due to dielectric<br />

absorption.<br />

When long lead lengths are required the following general practices can be used<br />

to minimize or measure settling errors:<br />

• Do not use wire with PVC-insulated conductors. PVC has a high dielectric,<br />

which extends input settling time.<br />

• Where possible, run excitation leads and signal leads in separate shields to<br />

minimize transients.<br />

• When measurement speed is not a prime consideration, additional time can be<br />

used to ensure ample settling time. The settling time required can be<br />

measured with the <strong>CR1000</strong>.<br />

•<br />

8.1.2.8.2 Measuring the Necessary Settling Time<br />

Settling time for a particular sensor and cable can be measured with the <strong>CR1000</strong>.<br />

Programming a series of measurements with increasing settling times will yield<br />

data that indicate at what settling time a further increase results in negligible<br />

change in the measured voltage. The programmed settling time at this point<br />

indicates the true settling time for the sensor and cable combination.<br />

CRBasic example Measuring Settling Time (p. 288) presents CRBasic code to help<br />

determine settling time for a pressure transducer utilizing a high-capacitance<br />

semi-conductor. The code consists of a series of full-bridge measurements<br />

(BrFull()) with increasing settling times. The pressure transducer is placed in<br />

287