Section 8. Operation<br />

duration. Consequently, noise at 1 / (integer multiples) of the integration duration<br />

is effectively rejected by an analog integrator. table CRBasic Measurement<br />

Integration Times and Codes (p. 284) lists three integration durations available in the<br />

<strong>CR1000</strong> and associated CRBasic codes. If reversing the differential inputs or<br />

reversing the excitation is specified, there are two separate integrations per<br />

measurement; if both reversals are specified, there are four separate integrations.<br />

Table 54. CRBasic Measurement Integration Times and Codes<br />

Integration Time (ms) CRBasic Code Comments<br />

250 µs 250 Fast integration<br />

16.667 ms _60Hz Filters 60-Hz noise<br />

20 ms _50Hz Filters 50-Hz noise<br />

8.1.2.7.1 ac Power Line Noise Rejection<br />

Grid or mains power (50 or 60 Hz, 230 or 120 Vac) can induce electrical noise at<br />

integer multiples of 50 or 60 Hz. Small analog voltage signals, such as<br />

thermocouples and pyranometers, are particularly susceptible. <strong>CR1000</strong> voltage<br />

measurements can be programmed to reject (filter) 50-Hz or 60-Hz related noise.<br />

ac Noise Rejection on Small Signals<br />

The <strong>CR1000</strong> rejects ac power line noise on all voltage ranges except mV5000 and<br />

mV2500 by integrating the measurement over exactly one ac cycle before A/D<br />

conversion as illustrated in table ac Noise Rejection on Small Signals (p. 284) and<br />

the full cycle technique of figure ac Power Line Noise Rejection Techniques (p.<br />

285).<br />

Table 55. ac Noise Rejection on Small Signals<br />

Applies to all analog input voltage ranges except mV2500 and mV5000.<br />

ac Power Line Frequency<br />

Measurement Integration<br />

Duration<br />

CRBasic Integration Code<br />

60 Hz 16.667 ms _60Hz<br />

50 Hz 20 ms _50Hz<br />

284