Section 8. Operation<br />

where<br />

Gain Error = ± (2500 * 0.0006)<br />

= ±1.5 mV<br />

and<br />

Offset Error = 1.5 • 667 µV + 1 µV = 1.00 mV<br />

Therefore,<br />

Error = Gain Error + Offset Error<br />

= ±1.5 mV + 1.00 µV<br />

= ±2.50 mV<br />

In contrast, the error for a 500‐mV input under the same constraints is<br />

±1.30 mV. The figure Voltage Measurement Accuracy (p. 279) illustrates<br />

the total error with respect to voltage measurements for the ±2500‐mV<br />

range.<br />

Figure 87: Voltage measurement accuracy (0° to 40°C)<br />

279