Section 7. Installation<br />

0.15°C over the -10 to 40°C temperature range. Because the measurement is<br />

ratiometric (R S /R f ), the properties of the 10-kΩ resistor do not affect the result.<br />

A terminal-input module (TIM) can be used to complete the circuit shown in<br />

figure PT100 in Four-Wire Half-Bridge (p. 259). Refer to the appendix Signal<br />

Conditioners (p. 561) for information concerning available TIM modules.<br />

Figure 79: PT100 in four-wire half-bridge<br />

CRBasic EXAMPLE. PT100 in Four-Wire Half-Bridge<br />

CRBasic Example 59.<br />

PT100 in Four‐Wire Half‐Bridge<br />

'See FIGURE. PT100 in Four-Wire Half-Bridge (p. 259) for wiring diagram<br />

Public Rs_Ro<br />

Public Deg_C<br />

BeginProg<br />

Scan(1,Sec,0,0)<br />

'BrHalf4W(Dest,Reps,Range1,Range2,DiffChan1,ExChan,MPS,Ex_mV,RevEx,RevDiff,<br />

' Settling, Integration,Mult,Offset)<br />

BrHalf4W(Rs_Ro,1,mV25,mV25,1,Vx1,1,2200,True,True,0,250,1.0111,0)<br />

'PRTCalc(Destination,Reps,Source,PRTType,Mult,Offset)<br />

PRTCalc(Deg_C,1,Rs_Ro,1,1.0,0) 'PRTType sets alpha<br />

NextScan<br />

EndProg<br />

7.8.18.2.3 PT100 in Three-Wire Half-Bridge<br />

Example shows:<br />

• How to measure a PRT in a three-wire half-bridge configuration.<br />

Advantages:<br />

• Uses half as many input channels as four-wire half-bridge.<br />

Disadvantages:<br />

• May not be as accurate as four-wire half-bridge.<br />

259